BD SYRINGE WITH NEEDLE
Report
- Report Number
- 3002682307-2019-00624
- Event Type
- Malfunction
- Date Received
- November 22, 2019
- Date of Event
- August 23, 2019
- Report Date
- December 16, 2019
- Manufacturer
- BECTON DICKINSON, S.A.
- Product Code
- FMF
- PMA / PMN Number
- N/A
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
INVESTIGATION SUMMARY: BD HAS NOT BEEN PROVIDED WITH PHOTOS OR SAMPLES FOR CATALOG 301948 LOT 1901183 TO INVESTIGATE FOR THIS RECORD. AS A RESULT, BD WAS UNABLE TO VERIFY THE REPORTED ISSUE OR DETERMINE A DEFINITIVE ROOT CAUSE. THE MATERIAL USED TO MANUFACTURE DISCARDIT SYRINGES HAS BEEN SELECTED AND TESTED TO RESIST NORMAL CONDITIONS OF USE. THE ASSEMBLING MACHINES HAVE AN ON-LINE DETECTION SYSTEM THAT INSPECTS 100% THE SYRINGES, REJECTING AUTOMATICALLY THE SYRINGES WITH BROKEN PARTS LIKE THE TIP OF THE BARREL. BASED ON THIS FACT, BD BELIEVES THAT THE SYRINGE TIP COULD BREAK BECAUSE OF A BLOCKAGE DURING MANUFACTURING IN THE PRIMARY PACKAGING MACHINE. CONSIDERING OUR IN-COMING AND IN-PROCESS INSPECTION AND SINCE THIS IS THE FIRST TIME THIS LOT IS REPORTED FOR THIS DEFECT, NO CORRECTIVE ACTIONS ARE REQUIRED AT THIS TIME.
IT HAS BEEN REPORTED THAT ONE BD¿ SYRINGE WITH NEEDLE HAS BEEN FOUND DEFORMED BEFORE USE. THE FOLLOWING HAS BEEN PROVIDED BY THE INITIAL REPORTER: THE TIP OF BARREL MISSING.
A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED, BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.
IT HAS BEEN REPORTED THAT ONE BD¿ SYRINGE WITH NEEDLE HAS BEEN FOUND DEFORMED BEFORE USE. THE FOLLOWING HAS BEEN PROVIDED BY THE INITIAL REPORTER: THE TIP OF BARREL MISSING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1160836 | BD SYRINGE WITH NEEDLE | SYRINGE WITH NEEDLE | FMF | BECTON DICKINSON, S.A. | 1901183 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |