LOGIQ P9
Report
- Report Number
- 9710090-2019-00004
- Event Type
- Death
- Date Received
- November 21, 2019
- Date of Event
- October 16, 2019
- Report Date
- December 19, 2019
- Manufacturer
- GE ULTRASOUND KOREA, LTD.
- Product Code
- IYN
- PMA / PMN Number
- K181783
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TN, US
- Reporter Occupation
- PHYSICIAN
Narratives
GEHC INVESTIGATION HAS COMPLETED. THE CUSTOMER/PHYSICIAN WAS CONTACTED, AND WHEN ASKED ABOUT THE ALLEGATION OF THE LOGIQ P9 MALFUNCTIONING AND CAUSING AND/OR CONTRIBUTING TO THE PATIENT DEATH THE PHYSICIAN SAID THE LOGIQ P9 DID NOT CAUSE OR CONTRIBUTE TO THE PATIENT DEATH. ADDITIONALLY, GEHC INVESTIGATION CONCLUDED THE LOGIQ P9 DID NOT MALFUNCTION, AND DUE TO THE LACK OF INFORMATION PROVIDED BY THE CUSTOMER THE ROOT CAUSE OF THE INABILITY TO SEND CINE IMAGES TO THEIR PACS CANNOT BE DETERMINED. GEHC BELIEVES THE MOST LIKELY CAUSE OF THE INABILITY TO SEND CINE IMAGES AT THE TIME WAS DUE TO EITHER A NETWORK AND/OR PACS ISSUE. THE CINE IMAGES WERE ABLE TO BE TRANSFERRED SUCCESSFULLY TO THEIR PACS FROM THE LOGIQ P9 ON A LATER DATE.
(B)(4). LEGAL MANUFACTURER: (B)(4). GE HEALTHCAREÂS SERVICE ENGINEER SERVICED THE LOGIQ P9 AND CONCLUDED THE LOGIQ P9 MAY HAVE MALFUNCTIONED. GE HEALTHCAREÂS INVESTIGATION IS ONGOING.
IT WAS REPORTED TO GE HEALTHCARE THAT A PHYSICIAN NEEDS A SERVICE ENGINEER TO ADJUST A SETTING ON THE LOGIQ P9 ULTRASOUND SYSTEM TO SEND CINE LOOPS TO THEIR NON-GE PACS BECAUSE THEY WILL NOT LOAD ONTO THEIR PACS SYSTEM. ADDITIONALLY, THE LOGIQ P9 WAS SLOW TO SPOOL IMAGES BEFORE SENDING TO PACS, AND THE SYSTEM LOCKED UP WHILE SCANNING. THE CONSEQUENCE IS THAT A PATIENT STUDY (DIAGNOSTIC PROCEDURE) COULD NOT BE LOADED AND READ ON THEIR PACS SYSTEM. PATIENT DIED BEFORE THEIR STUDY COULD BE READ.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1149830 | LOGIQ P9 | ULTRASONIC IMAGING SYSTEM | IYN | GE ULTRASOUND KOREA, LTD. | H42872LB | LP9003946 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Life Threatening |