FDA Adverse Event Malfunction Summary report: N

SYRINGE S2 2ML 23GA 1-1/4IN BD

MDR report key: 9355600 · Received November 21, 2019

Report

Report Number
3002682307-2019-00619
Event Type
Malfunction
Date Received
November 21, 2019
Date of Event
April 30, 2019
Report Date
December 16, 2019
Manufacturer
BECTON DICKINSON, S.A.
Product Code
FMF
PMA / PMN Number
N/A
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

H.6. INVESTIGATION SUMMARY: BD HAS NOT BEEN PROVIDED WITH PHOTOS OR SAMPLES FOR CATALOG 301940 LOT 1803120 TO INVESTIGATE FOR THIS RECORD. AS A RESULT, BD WAS UNABLE TO VERIFY THE REPORTED ISSUE OR DETERMINE A DEFINITIVE ROOT CAUSE. BD CONCLUDES THAT THE CAUSE OF THE PROBLEM COULD BE PRODUCED BECAUSE OF A DAMAGE IN THE PLUNGER LIP. THIS COULD BE PRODUCED DURING THE HANDLING OF THE PRODUCT THROUGH THE MANUFACTURING PROCESS OR IN THE PLUNGER ASSEMBLY MACHINE. CONSIDERING OUR IN-COMING AND IN-PROCESS INSPECTION AND SINCE THIS IS THE FIRST TIME THIS LOT IS REPORTED FOR THIS DEFECT, NO CORRECTIVE ACTIONS ARE REQUIRED AT THIS TIME. H3 OTHER TEXT : SEE SECTION H.10.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE SYRINGE S2 2ML 23GA 1-1/4IN BD CHINA EXPERIENCED LEAKAGE WHICH WAS NOTED PRIOR TO USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: FOUND LEAK FROM THE PLUNGER ROD. THE LOT # PROVIDED BY CUTOMER IS 1803120.

Additional Manufacturer Narrative · 1

A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE SYRINGE S2 2ML 23GA 1-1/4IN BD (B)(4) EXPERIENCED LEAKAGE WHICH WAS NOTED PRIOR TO USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: FOUND LEAK FROM THE PLUNGER ROD. THE LOT # PROVIDED BY CUSTOMER IS 1803120.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1154166 SYRINGE S2 2ML 23GA 1-1/4IN BD SYRINGE FMF BECTON DICKINSON, S.A. 1803120

Patients

Seq Age Sex Outcome Treatment
1 Other