BD ULTRASAFE X100L PNG RASPBERRY NVS STEIN
Report
- Report Number
- 3009081593-2019-00257
- Event Type
- Malfunction
- Date Received
- November 21, 2019
- Date of Event
- September 22, 2019
- Report Date
- November 22, 2019
- Manufacturer
- BECTON DICKINSON HUNGARY KFT (BD)
- Product Code
- MEG
- PMA / PMN Number
- SEE H.10
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SZ
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
INVESTIGATION: NEITHER PHOTO NOR SAMPLE WAS PROVIDED TO BD MEDICAL ¿ PHARMACEUTICAL SYSTEM (BDM-PS) FOR ANALYSIS. BDM-PS PERFORMED A BATCH HISTORY RECORD¿S REVIEW (BHR) INCLUDING A REVIEW OF ALL DATA COLLECTED DURING IN PROCESS AND QUALITY INSPECTIONS. THE BATCHES INVOLVED IN THIS COMPLAINT MEET ALL ACCEPTABLE QUALITY LEVELS (AQL¿S), WERE MANUFACTURED AND RELEASED ACCORDING TO APPLICABLE PROCEDURES AND SPECIFICATIONS. BASED ON INVESTIGATION CONCLUSION A SYRINGE CAN ONLY BECOME DETACHED IF SYRINGE CAPTURE FEATURES OR THE FLANGE OF THE SYRINGE GET DAMAGED/BROKEN OR IF THE SYRINGE RECEIVES AN IMPACT AFTER SYRINGE INSERTION WHICH CAUSES IT TO UNCLIP FROM THE DEVICE. THEREFORE, THE SYRINGE MOST LIKELY BECAME DETACHED AS IT WAS NOT CLIPPED INTO THE DEVICE PROPERLY OR IT RECEIVED AN EXTERNAL IMPACT AFTER SYRINGE INSERTION WHICH CAUSED IT TO UNCLIP FROM THE DEVICE. NONE OF THESE CAUSES ARE RELATED TO BD PROCESS.
IT WAS REPORTED THAT THE PREFILLED SYRINGE FELL OUT OF THE NEEDLE SAFETY DEVICE PRIOR TO USE WITH A BD ULTRASAFE¿ X100L PNG RASPBERRY NVS STEIN. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "I PICKED IT UP TO BRING TO THE PATIENT AND THE SYRINGE FELL OUT AND DROPPED ON THE FLOOR. I DIDN'T GET A CHANCE TO EVEN TAKE OFF THE NEEDLE CAP OR USE THE SYRINGE ON THE PATIENT."
"MULTIPLE LOT NUMBERS: THERE WERE MULTIPLE POSSIBLE LOT NUMBERS REPORTED TO BE INVOLVED. THE INFORMATION FOR EACH LOT NUMBER IS AS FOLLOWS: MEDICAL DEVICE LOT #: 8068790, MEDICAL DEVICE EXPIRATION DATE: 2023-02-28, DEVICE MANUFACTURE DATE: 2018-03-09. MEDICAL DEVICE LOT #: 8071890, MEDICAL DEVICE EXPIRATION DATE: 2023-02-28, DEVICE MANUFACTURE DATE: 2018-03-12." PMA / 510(K)#: K011369, K122558. INITIAL REPORTER PHONE #: UNKNOWN. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
IT WAS REPORTED THAT THE PREFILLED SYRINGE FELL OUT OF THE NEEDLE SAFETY DEVICE PRIOR TO USE WITH A BD ULTRASAFE¿ X100L PNG RASPBERRY NVS STEIN. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "I PICKED IT UP TO BRING TO THE PATIENT AND THE SYRINGE FELL OUT AND DROPPED ON THE FLOOR. I DIDN'T GET A CHANCE TO EVEN TAKE OFF THE NEEDLE CAP OR USE THE SYRINGE ON THE PATIENT."
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1154579 | BD ULTRASAFE X100L PNG RASPBERRY NVS STEIN | PISTON SYRINGE | MEG | BECTON DICKINSON HUNGARY KFT (BD) | SEE H.10 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |