FDA Adverse Event Malfunction Summary report: N

BD PLASTIPAK CONCENTRIC LUER LOCK SYRINGE

MDR report key: 9355190 · Received November 21, 2019

Report

Report Number
3003152976-2019-00789
Event Type
Malfunction
Date Received
November 21, 2019
Date of Event
November 6, 2019
Report Date
December 4, 2019
Manufacturer
BECTON DICKINSON, S.A.
Product Code
FMF
PMA / PMN Number
N/A
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

ONE PHOTO WAS PROVIDED TO OUR QUALITY TEAM FOR INVESTIGATION. UPON VISUAL INSPECTION OF THE PHOTO, A LEAK WAS OBSERVED BETWEEN THE STOPPER RIBS. A DEVICE HISTORY REVIEW WAS PERFORMED FOR THE REPORTED LOTS 1905234 AND 1907257, NO DEVIATIONS OR NON-CONFORMANCES WERE IDENTIFIED DURING THE MANUFACTURING PROCESS THAT COULD HAVE CONTRIBUTED TO THIS ISSUE. TEN RETAINED SAMPLES OF LOT 1905234 AND AN ADDITIONAL TEN SAMPLES OF LOT 1907257 WERE USED TO CONDUCT A LEAKAGE TEST. THE PRODUCT WAS VISUALLY INSPECTED, NO DEFECTS OR DAMAGE WAS NOTED, AND NO LEAK WAS IDENTIFIED. BASED ON THE AVAILABLE INFORMATION WE ARE NOT ABLE TO DETERMINE A ROOT CAUSE AT THIS TIME. COMPLAINTS RECEIVED FOR THIS DEFECT AND DEVICE WILL BE MONITORED BY OUR QUALITY TEAM FOR SIGNS OF EMERGING TRENDS.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE BD PLASTIPAK¿ CONCENTRIC LUER LOCK SYRINGES LEAKED CHEMOTHERAPY MEDICATION DURING USE. LOT #'S 1905234 AND 1907257 WERE REPORTED TO HAVE BEEN INVOLVED IN THIS EVENT, BUT IT IS UNKNOWN HOW MANY OCCURRENCES HAPPENED WITHIN EACH LOT. THIS COMPLAINT WAS CREATED TO CAPTURE THE 1ST OF 2 RELATED INCIDENTS. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM FRENCH TO ENGLISH: "WE HAD LEAKAGE PROBLEMS ON AT LEAST 2 BATCHES OF 50ML SYRINGES THAT WE USED FOR THE PREPARATION AND ADMINISTRATION OF CHEMOTHERAPY. PROBLEM OF COSTLY PRODUCT LOST FOR MOST AND ESPECIALLY EXPOSURE OF PERSONNEL DURING PREPARATION WITH INCREASED RISK OF CONTAMINATION. PROBLEM ENCOUNTERED BY SEVERAL USERS WITHOUT CHANGING RECENT PRACTICES. THE LOTS CONCERNED TO DATE ARE 1905234 AND 1907257."

Additional Manufacturer Narrative · 1

DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. THERE WERE MULTIPLE LOT NUMBERS REPORTED TO BE INVOLVED. THE INFORMATION FOR EACH LOT NUMBER IS AS FOLLOWS: MEDICAL DEVICE LOT #: 1905234, MEDICAL DEVICE EXPIRATION DATE: 2024-04-30, DEVICE MANUFACTURE DATE: 2019-05-16, MEDICAL DEVICE LOT #: 1907257, MEDICAL DEVICE EXPIRATION DATE: 2024-06-30, DEVICE MANUFACTURE DATE: 2019-07-31. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE BD PLASTIPAK¿ CONCENTRIC LUER LOCK SYRINGES LEAKED CHEMOTHERAPY MEDICATION DURING USE. LOT #'S 1905234 AND 1907257 WERE REPORTED TO HAVE BEEN INVOLVED IN THIS EVENT, BUT IT IS UNKNOWN HOW MANY OCCURRENCES HAPPENED WITHIN EACH LOT. THIS COMPLAINT WAS CREATED TO CAPTURE THE 1ST OF 2 RELATED INCIDENTS. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM FRENCH TO ENGLISH: "WE HAD LEAKAGE PROBLEMS ON AT LEAST 2 BATCHES OF 50 ML SYRINGES THAT WE USED FOR THE PREPARATION AND ADMINISTRATION OF CHEMOTHERAPY. PROBLEM OF COSTLY PRODUCT LOST FOR MOST AND ESPECIALLY EXPOSURE OF PERSONNEL DURING PREPARATION WITH INCREASED RISK. OF CONTAMINATION. PROBLEM ENCOUNTERED BY SEVERAL USERS WITHOUT CHANGING RECENT PRACTICES. THE LOTS CONCERNED TO DATE ARE 1905234 AND 1907257."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1154183 BD PLASTIPAK CONCENTRIC LUER LOCK SYRINGE PISTON SYRINGE FMF BECTON DICKINSON, S.A. SEE SECTION H.10.

Patients

Seq Age Sex Outcome Treatment
1 Other