BD PLASTIPAK CONCENTRIC LUER LOCK SYRINGE
Report
- Report Number
- 3003152976-2019-00789
- Event Type
- Malfunction
- Date Received
- November 21, 2019
- Date of Event
- November 6, 2019
- Report Date
- December 4, 2019
- Manufacturer
- BECTON DICKINSON, S.A.
- Product Code
- FMF
- PMA / PMN Number
- N/A
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
ONE PHOTO WAS PROVIDED TO OUR QUALITY TEAM FOR INVESTIGATION. UPON VISUAL INSPECTION OF THE PHOTO, A LEAK WAS OBSERVED BETWEEN THE STOPPER RIBS. A DEVICE HISTORY REVIEW WAS PERFORMED FOR THE REPORTED LOTS 1905234 AND 1907257, NO DEVIATIONS OR NON-CONFORMANCES WERE IDENTIFIED DURING THE MANUFACTURING PROCESS THAT COULD HAVE CONTRIBUTED TO THIS ISSUE. TEN RETAINED SAMPLES OF LOT 1905234 AND AN ADDITIONAL TEN SAMPLES OF LOT 1907257 WERE USED TO CONDUCT A LEAKAGE TEST. THE PRODUCT WAS VISUALLY INSPECTED, NO DEFECTS OR DAMAGE WAS NOTED, AND NO LEAK WAS IDENTIFIED. BASED ON THE AVAILABLE INFORMATION WE ARE NOT ABLE TO DETERMINE A ROOT CAUSE AT THIS TIME. COMPLAINTS RECEIVED FOR THIS DEFECT AND DEVICE WILL BE MONITORED BY OUR QUALITY TEAM FOR SIGNS OF EMERGING TRENDS.
IT WAS REPORTED THAT THE BD PLASTIPAK¿ CONCENTRIC LUER LOCK SYRINGES LEAKED CHEMOTHERAPY MEDICATION DURING USE. LOT #'S 1905234 AND 1907257 WERE REPORTED TO HAVE BEEN INVOLVED IN THIS EVENT, BUT IT IS UNKNOWN HOW MANY OCCURRENCES HAPPENED WITHIN EACH LOT. THIS COMPLAINT WAS CREATED TO CAPTURE THE 1ST OF 2 RELATED INCIDENTS. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM FRENCH TO ENGLISH: "WE HAD LEAKAGE PROBLEMS ON AT LEAST 2 BATCHES OF 50ML SYRINGES THAT WE USED FOR THE PREPARATION AND ADMINISTRATION OF CHEMOTHERAPY. PROBLEM OF COSTLY PRODUCT LOST FOR MOST AND ESPECIALLY EXPOSURE OF PERSONNEL DURING PREPARATION WITH INCREASED RISK OF CONTAMINATION. PROBLEM ENCOUNTERED BY SEVERAL USERS WITHOUT CHANGING RECENT PRACTICES. THE LOTS CONCERNED TO DATE ARE 1905234 AND 1907257."
DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. THERE WERE MULTIPLE LOT NUMBERS REPORTED TO BE INVOLVED. THE INFORMATION FOR EACH LOT NUMBER IS AS FOLLOWS: MEDICAL DEVICE LOT #: 1905234, MEDICAL DEVICE EXPIRATION DATE: 2024-04-30, DEVICE MANUFACTURE DATE: 2019-05-16, MEDICAL DEVICE LOT #: 1907257, MEDICAL DEVICE EXPIRATION DATE: 2024-06-30, DEVICE MANUFACTURE DATE: 2019-07-31. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
IT WAS REPORTED THAT THE BD PLASTIPAK¿ CONCENTRIC LUER LOCK SYRINGES LEAKED CHEMOTHERAPY MEDICATION DURING USE. LOT #'S 1905234 AND 1907257 WERE REPORTED TO HAVE BEEN INVOLVED IN THIS EVENT, BUT IT IS UNKNOWN HOW MANY OCCURRENCES HAPPENED WITHIN EACH LOT. THIS COMPLAINT WAS CREATED TO CAPTURE THE 1ST OF 2 RELATED INCIDENTS. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM FRENCH TO ENGLISH: "WE HAD LEAKAGE PROBLEMS ON AT LEAST 2 BATCHES OF 50 ML SYRINGES THAT WE USED FOR THE PREPARATION AND ADMINISTRATION OF CHEMOTHERAPY. PROBLEM OF COSTLY PRODUCT LOST FOR MOST AND ESPECIALLY EXPOSURE OF PERSONNEL DURING PREPARATION WITH INCREASED RISK. OF CONTAMINATION. PROBLEM ENCOUNTERED BY SEVERAL USERS WITHOUT CHANGING RECENT PRACTICES. THE LOTS CONCERNED TO DATE ARE 1905234 AND 1907257."
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1154183 | BD PLASTIPAK CONCENTRIC LUER LOCK SYRINGE | PISTON SYRINGE | FMF | BECTON DICKINSON, S.A. | SEE SECTION H.10. |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |