JTS DRIVE UNIT
Report
- Report Number
- 3004105610-2019-00122
- Event Type
- Malfunction
- Date Received
- November 21, 2019
- Date of Event
- September 6, 2019
- Report Date
- January 23, 2020
- Manufacturer
- STANMORE IMPLANTS WORLDWIDE
- Product Code
- KRO
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IT
- Reporter Occupation
- PHYSICIAN
Narratives
ON INITIAL RECEIPT OF THE REPORTED EVENT, THE EVENT DESCRIPTION WAS REVIEWED AND WITH THE LIMITED INFORMATION AVAILABLE AT THAT TIME, A DECISION WAS MADE TO REPORT THE EVENT. THE INVESTIGATION CONFIRMED THAT THIS UNIT WAS FUNCTIONAL AND DID NOT CAUSE OR CONTRIBUTE TO THE REPORTED EVENT. THE EVENT IS THEREFORE, NOT REPORTABLE.
THIS PI IS CREATED FOR THE FIRST ATTEMPT LAST 25 OF AUGUST. IT IS REPORTED " FIRST ATTEMPT HAS BEEN PERFORMED WITH NO SUCCESS LAST 25 OF AUGUST WITH POWER UNIT SET ON 0. SECOND ATTEMPT TODAY." ADDITIONAL NOTES STATE "CENTER IS EXPOSED TO JTS SYSTEM, NEVER GOT PROBLEMS BEFORE. SAME SKILLED PHYSICIAN AS IN THE PAST. IT HAS BEEN TRIED SEVERAL TIME THE UNLOCKING PROCEDURE TO SWITCH FROM B TO A. POWER UNIT SET ON 3 AND 4. IMPLANT SOUNDED AS LOCKED, SOUNDED WORKING GOING REVERSE (B) AND THEN GOING AHEAD (A) BUT JUST FOR A FEW SECONDS. "
PLEASE NOTE THAT THIS EVENT IS RELATED TO A JTS DRIVE UNIT WHICH IS USED IN CONJUNCTION WITH THE JTS NON-INVASIVE EXTENDIBLE IMPLANT (K092138). AN INVESTIGATION IS BEING PERFORMED IN AN ATTEMPT TO IDENTIFY THE CAUSE OF THE EVENT. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE IT WILL BE REPORTED IN A SUPPLEMENTAL REPORT.
THIS PI IS CREATED FOR THE FIRST ATTEMPT LAST 25 OF AUGUST. IT IS REPORTED " FIRST ATTEMPT HAS BEEN PERFORMED WITH NO SUCCESS LAST 25 OF AUGUST WITH POWER UNIT SET ON 0. SECOND ATTEMPT TODAY." ADDITIONAL NOTES STATE "CENTER IS EXPOSED TO JTS SYSTEM, NEVER GOT PROBLEMS BEFORE. SAME SKILLED PHYSICIAN AS IN THE PAST. IT HAS BEEN TRIED SEVERAL TIME THE UNLOCKING PROCEDURE TO SWITCH FROM B TO A. POWER UNIT SET ON 3 AND 4. IMPLANT SOUNDED AS LOCKED, SOUNDED WORKING GOING REVERSE (B) AND THEN GOING AHEAD (A) BUT JUST FOR A FEW SECONDS."
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1150404 | JTS DRIVE UNIT | PROSTHESIS, KNEE, FEMOROTIBIAL, CONSTRAINED, CEMENTED, METAL/POLYMER | KRO | STANMORE IMPLANTS WORLDWIDE | 907-110 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 14 YR | Other |