FDA Adverse Event Injury Summary report: N

CORTSCR Ø2.7 SELF-TAP L22 SST

MDR report key: 9350036 · Received November 20, 2019

Report

Report Number
8030965-2019-70430
Event Type
Injury
Date Received
November 20, 2019
Report Date
October 22, 2019
Manufacturer
OBERDORF SYNTHES PRODUKTIONS GMBH
Product Code
HWC
UDI-DI
07611819936331
PMA / PMN Number
K113364
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 0

DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. H10 ADDITIONAL NARRATIVE: H3, H6: INVESTIGATION SUMMARY. DEVICE HISTORY LOT: STERILE, PART: 202.882S, LOT: 9132924, MANUFACTURING SITE: SELZACH, SUPPLIER: IL-MEDTEC AG, RELEASE TO WAREHOUSE DATE: SEP 08, 2014, EXPIRY DATE: AUG 01, 2024. A MANUFACTURING RECORD EVALUATION WAS PERFORMED FOR THE FINISHED DEVICE LOT NUMBER, AND NO NON-CONFORMANCES WERE IDENTIFIED. NON-STERILE, PART: 202.882, LOT: 9063568, MANUFACTURING SITE: MEZZOVICO, RELEASE TO WAREHOUSE DATE: JUL 16, 2014. A MANUFACTURING RECORD EVALUATION WAS PERFORMED FOR THE FINISHED DEVICE LOT NUMBER, AND NO NON-CONFORMANCES WERE IDENTIFIED. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 1

DEVICE AVAILABLE FOR EVALUATION: COMPLAINANT PART IS NOT EXPECTED TO BE RETURNED FOR MANUFACTURER REVIEW/INVESTIGATION. THE INVESTIGATION COULD NOT BE COMPLETED; NO CONCLUSION COULD BE DRAWN, AS NO PRODUCT WAS RECEIVED. BASED ON THE INFORMATION AVAILABLE, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. THIS COMPLAINT WILL BE ACCOUNTED FOR AND MONITORED VIA POST MARKET SURVEILLANCE ACTIVITIES. IF ADDITIONAL INFORMATION IS MADE AVAILABLE, THE INVESTIGATION WILL BE UPDATED AS APPLICABLE. (B)(4). DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 1

DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED PATIENT UNDERWENT SURGERY FOR LEFT FIRST METATARSOPHALANGEAL JOINT FUSION ON (B)(6) 2018. A STANDARD TECHNIQUE WAS USED. THE FUSION WAS STABILIZED WITH A 42MM LEFT ZERO DEGREE 1.MTP FUSION PLATE REF 20.211.231S LOT H826318 (DEPUY SYNTHES). ON ABOUT (B)(6) 2019 THE PATIENT CONTACTED THE TREATMENT CENTRE WHERE I AM INFORMED THAT SUPERFICIAL INFECTION WAS DIAGNOSED AND FLUCLOXACILLIN STARTED. THE COMPLETE ABSENCE IN THE MEDICAL RECORDS OF THIS ENCOUNTER IS NOTED. I SAW THE PATIENT IN OUTPATIENT CLINIC ON (B)(6) 2019 FOR A PLANNED FOLLOW UP APPOINTMENT AND NOTED THE JOINT RED AND MORE SWOLLEN THAN IS USUALLY TO BE EXPECTED AND AN X-RAY TAKEN THAT DAY SHOWED THE FUSION PLATE TO HAVE BROKEN AND THE JOINT NOT FUSED WITH SOFT TISSUE SWELLING. ACTIVE MONITORING OF THE PATIENT, WITH POTENTIAL TO REMOVE METAL AND CARRY OUT REVISION SURGERY. THIS COMPLAINT INVOLVES SEVEN (7) DEVICES. THIS IS 4 OF 5 FOR REPORT (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1143588 CORTSCR Ø2.7 SELF-TAP L22 SST SCREW, FIXATION, BONE HWC OBERDORF SYNTHES PRODUKTIONS GMBH 9132924 07611819936331

Patients

Seq Age Sex Outcome Treatment
1 74 YR Required Intervention