CATALYS SYSTEM
Report
- Report Number
- 3005675890-2019-00099
- Event Type
- Malfunction
- Date Received
- November 20, 2019
- Date of Event
- October 28, 2019
- Report Date
- November 5, 2020
- Manufacturer
- JOHNSON & JOHNSON SURGICAL VISION, INC.
- Product Code
- OOE
- UDI-DI
- 05050474609792
- PMA / PMN Number
- K121091
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- OTHER
Narratives
THIS CORRECTION MDR IS BEING SUBMITTED TO ADDRESS INACCURATE INFORMATION IN SECTION F ON THE XML VERSION OF THE PREVIOUS SUBMISSION CAUSED BY A SOFTWARE GLITCH IN THE TRANSMISSION PROCESS. THE INTERNAL NON-CONFORMANCE NUMBERS FOR THIS SOFTWARE ISSUE ARE NR-0148919 AND CAPA-010215.
AMO¿S INVESTIGATION SUBSEQUENTLY IDENTIFIED THAT THE CATALYS SYSTEM MAY HAVE A REMOTE POTENTIAL EVENT WHERE LOSS OF SUCTION DURING LASER FIRING COULD CREATE A HAZARDOUS SITUATION THAT MAY RESULT IN SCORING OF THE POSTERIOR CORNEAL SURFACE. THEREFORE JOHNSON & JOHNSON SURGICAL VISION, INC HAS ISSUED AN ADVISORY NOTICE TO REINFORCE INSTRUCTIONS PROVIDED IN PRODUCT TRAINING AND IN THE OPERATOR¿S MANUAL REGARDING SUCTION LOSS DURING A PROCEDURE. A REVIEW OF RECORDS RELATED TO THE DEVICE INCLUDING LABELING, COMPLAINT TRENDING, AND RISK DOCUMENTATION WILL BE PERFORMED. UPON COMPLETION OF THIS REVIEW, IF THERE IS ANY FURTHER RELEVANT INFORMATION OBTAINED, A SUPPLEMENTAL MEDWATCH WILL BE FILED. ALL PERTINENT INFORMATION AVAILABLE TO JOHNSON & JOHNSON SURGICAL VISION, INC HAS BEEN SUBMITTED.
JOHNSON & JOHNSON SURGICAL VISION, INC. APPLICATION SUPPORT MANAGER (ASM) REPORTED THAT SUCTION WAS LOST DURING CAPSULOTOMY TREATMENT (DURING LASER FIRING) AND WAS DUE TO EXCESSIVE PATIENT MOVEMENT. THE SURGEON COMPLETED THE CASE MANUALLY IN THE OPERATING ROOM (OR) WITHOUT ISSUE AND THE PATIENT DID FINE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1144755 | CATALYS SYSTEM | CATALYS | OOE | JOHNSON & JOHNSON SURGICAL VISION, INC. | CATALYS-U | 05050474609792 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |