FDA Adverse Event Death Summary report: N

ICU MONITORING SYSTEM

MDR report key: 93451 · Received May 21, 1997

Report

Report Number
93451
Event Type
Death
Date Received
May 21, 1997
Date of Event
February 27, 1997
Report Date
May 15, 1997
Manufacturer
HEWLETT PACKARD WEST SALES REGION
Product Code
DRT
Adverse Event
Yes
Product Problem
Yes
Report Source
User Facility report
Reporter Location
MO, US
Reporter Occupation
RISK MANAGER

Narratives

Description of Event or Problem · 1

ON FEBRUARY 27, 1997, A 45 YEAR OLD MALE PT WAS ADMITTED VIA THE E.R. TO THE ICU AT ABOUT 1230 HOURS. ADMITTING DIAGNOSIS INCLUDES CHF, PNEUMONIA, SEIZURES, AND ALCOHOLIC CIRRHOSIS. AT APPROX 1858 HOURS PT WAS COINCIDENTALLY FOUND BY ICU STAFF IN A CODE CONDITION. A CODE WAS CALLED, STAFF RESPONDED, THE PT SURVIVED UNTIL APPROX 0900 HOURS ON FEBRUARY 28, 1997, AT WHICH TIME DEATH WAS PRONOUNCED. INVESTIGATION INTO THE MATTER ASCERTAINED THE ECG ALARM SWITCH LOCATED ON THE BEDSIDE MONITOR (HEWLETT PACKARD MODEL 78532B) WAS IN THE OFF POSITION. CONSEQUENTLY APPROPRIATE INFO COULD NOT BE TRANSMITTED TO THE CENTRAL MONITORING SYSTEM LOCATED AT THE NURSING STATION. FURTHERMORE THE CENTRAL SYSTEM COULD NOT GO INTO THE ALARM MODE THEREBY ALERTING NURSING PERSONNEL THAT THE PT WAS IN A CODE CONDITION. THIS EVENT IS VIEWED REPORTABLE FROM A) USER ERROR, AND B) EQUIPMENT DESIGN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ICU MONITORING SYSTEM BEDSIDE MONITOR DRT HEWLETT PACKARD WEST SALES REGION 78532B *

Patients

Seq Age Sex Outcome Treatment
1 45 YR Death