FDA Adverse Event Malfunction Summary report: N

ADVIA CHEMISTRY XPT

MDR report key: 9337567 · Received November 18, 2019

Report

Report Number
2432235-2019-00421
Event Type
Malfunction
Date Received
November 18, 2019
Date of Event
October 29, 2019
Report Date
December 16, 2019
Manufacturer
SIEMENS HEALTHCARE DIAGNOSTICS, INC.
Product Code
JJE
UDI-DI
00630414595467
PMA / PMN Number
K990346
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FI
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

SIEMENS FILED AN INITIAL MDR 2432235-2019-00421 ON 18-NOV-2019 FOR A DISCORDANT, FALSELY ELEVATED TOBRAMYCIN_2 PATIENT RESULT. ADDITIONAL INFORMATION (21-NOV-2019): SIEMENS WAS INFORMED THAT THE PHYSICIAN WAS EXPECTING A RESULT OF < 0.5 UG/ML ADDITIONAL INFORMATION (04-DEC-2019): SIEMENS HAS REVIEWED THE INFORMATION PROVIDED. THE CUSTOMER INITIALLY RECOVERED A RESULT (K) FLAG WHICH IS CONSISTENT WITH A RESULT BELOW THE CALIBRATOR LOW RANGE (<0.4 UG/ML). THE ORIGINAL SAMPLE WAS REPEATED AND RECOVERED A DISCORDANT VALUE OF 2.5 UG/ML. ADDITIONAL REPEATS RECOVERED VALUES CONSISTENT WITH THE ORIGINAL RESULT. CALIBRATION WAS WITHIN CONFORMANCE AND QUALITY CONTROL (QC) RESULTS RECOVERED WITHIN THE CUSTOMER ESTABLISHED RANGES. A REVIEWED OF THE INSTRUMENT PRINTOUTS AND INSTRUMENT FILTER-DATA ASSOCIATED WITH THE DISCORDANT RESULT INDICATED ELEVATED ABSORBANCE READINGS FOLLOWING REAGENT 1 AND SAMPLE ADDITION TO THE REACTION CUVETTE. POTENTIAL CAUSES MAY INCLUDE AN ATYPICAL REAGENT 1 OR SAMPLE ADDITION AS WELL AS PRE-ANALYTICAL SAMPLE HANDLING. A SERVICE ENGINEER WENT ON-SITE AND COMPLETED DILUTION PROBE PIPETTE (DDP) AND SAMPLE PROBE PIPETTE (SSP) ALIGNMENTS WHICH ARE A PART OF NORMAL TROUBLESHOOTING. THE CUSTOMER HAS NOT REPORTED ANY ADDITIONAL DISCORDANT TOBRAMYCIN_2 RESULTS. THE CUSTOMER IS OPERATIONAL. NO PRODUCT NON-CONFORMANCE WAS IDENTIFIED. THE INSTRUMENT IS PERFORMING WITHIN SPECIFICATIONS. NO FURTHER EVALUATION OF THE DEVICE IS REQUIRED.

Additional Manufacturer Narrative · 1

THE CUSTOMER CONTACTED THE SIEMENS CUSTOMER CARE CENTER (CCC) TO REPORT A FALSELY ELEVATED TOBRAMYCIN_2 PATIENT RESULT. A SIEMENS CUSTOMER SERVICE ENGINEER (CSE) HAS VERIFIED THE DILUTION PROBE PIPETTE (DDP) AND SAMPLE PROBE PIPETTE (SSP) ALIGNMENTS TO THE CUVETTES. QUALITY CONTROL RESULTS WERE WITHIN ACCEPTABLE RANGES AT THE TIME OF THE EVENT. SIEMENS IS INVESTIGATING.

Description of Event or Problem · 1

A DISCORDANT, FALSELY ELEVATED TOBRAMYCIN_2 RESULT WAS OBTAINED UPON REPEAT ON A NEWBORN SAMPLE ON AN ADVIA CHEMISTRY XPT INSTRUMENT. THE PATIENT'S INITIAL TOBRAMYCIN RESULTED LOW (BELOW THE ASSAY MEASURING RANGE), SO THE CUSTOMER REPEATED THE SAME SAMPLE TWO MORE TIMES, RESULTING LOW. THE LOWER TOBRAMYCIN_2 RESULT WAS REPORTED TO THE PHYSICIAN(S) AS THE CORRECT RESULT. THERE ARE NO KNOWN REPORTS OF ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT, FALSELY ELEVATED TOBRAMYCIN_2 RESULT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1128501 ADVIA CHEMISTRY XPT IMMUNOASSAY ANALYZER JJE SIEMENS HEALTHCARE DIAGNOSTICS, INC. ADVIA CHEMISTRY XPT 00630414595467

Patients

Seq Age Sex Outcome Treatment
1 5 DA