FDA Adverse Event Injury Summary report: N

COMPACT INTUITIV CONSOLE

MDR report key: 9337408 · Received November 18, 2019

Report

Report Number
3006695864-2019-00905
Event Type
Injury
Date Received
November 18, 2019
Date of Event
October 14, 2019
Report Date
November 5, 2020
Manufacturer
JOHNSON & JOHNSON SURGICAL VISION, INC.
Product Code
HQC
UDI-DI
05050474573628
PMA / PMN Number
K060366
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
OR, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 0

CORRECTION: IN INITIAL REPORT, THE MANUFACTURING DATE WAS REPORTED AS 8/22/2017. THE CORRECT DATE IS 7/11/2017. ALL PERTINENT INFORMATION AVAILABLE TO JOHNSON & JOHNSON SURGICAL VISION, INC HAS BEEN SUBMITTED.

Additional Manufacturer Narrative · 0

THIS CORRECTION MDR IS BEING SUBMITTED TO ADDRESS INACCURATE INFORMATION IN SECTION F ON THE XML VERSION OF THE PREVIOUS SUBMISSION CAUSED BY A SOFTWARE GLITCH IN THE TRANSMISSION PROCESS. THE INTERNAL NON-CONFORMANCE NUMBERS FOR THIS SOFTWARE ISSUE ARE NR-0148919 AND CAPA: 010215.

Additional Manufacturer Narrative · 1

THE SYSTEM WAS EVALUATED BY A FIELD SERVICE ENGINEER. THE FIELD SERVICE FOUND NO ISSUES WITH THE UNIT. A FIELD SERVICE CHECKLIST WAS PERFORMED. THE UNIT COMPLIED WITH ALL FACTORY SETTINGS. A REVIEW OF RECORDS RELATED TO THE DEVICE INCLUDING LABELING, COMPLAINT TRENDING, AND RISK DOCUMENTATION WILL BE PERFORMED. UPON COMPLETION OF THIS REVIEW, IF THERE IS ANY FURTHER RELEVANT INFORMATION OBTAINED, A SUPPLEMENTAL MEDWATCH WILL BE FILED. ALL PERTINENT INFORMATION AVAILABLE TO JOHNSON & JOHNSON SURGICAL VISION, INC. HAS BEEN SUBMITTED.

Description of Event or Problem · 1

THE SURGEON REPORTED THAT DURING CATARACT SURGERY ON (B)(6) 2019, ONCE HE APPLIED THE PHACO TIP TO THE PATIENT'S LEFT EYE HE IMMEDIATELY SAW A WHITE PLUME APPEAR AND REMOVED THE TIP. THE PATIENT REQUIRED A SINGLE FIGURE 8 SUTURE FOLLOWING THE CORNEAL BURN. THE SURGEON REPORTED THAT THE EVENT MAY HAVE BEEN DUE TO PHYSICIAN TECHNIQUE, BUT HE COULD NOT CONFIRM THIS. ON (B)(6) 2019, A SECONDARY PROCEDURE WAS PERFORMED ON THE LEFT EYE BY APPLYING AN ADDITIONAL SINGLE FIGURE 8 SUTURE OVER THE INCISION. THE PATIENT IS STILL HEALING. THIS REPORT IS FOR THE COMPACT INTUITIV PHACO SYSTEM. AN ADDITIONAL REPORT IS BEING SUBMITTED FOR THE 21 GAUGE LAMINAR PHACO TIP.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1129121 COMPACT INTUITIV CONSOLE COMPACT INTUITIV HQC JOHNSON & JOHNSON SURGICAL VISION, INC. SCP680300 05050474573628

Patients

Seq Age Sex Outcome Treatment
1 64 YR Required Intervention PHACO TIP| PHACO TIP