FDA Adverse Event
Malfunction
Summary report: N
INRATIO
MDR report key: 933454
·
Received October 22, 2007
Report
- Report Number
- 2954730-2007-00516
- Event Type
- Malfunction
- Date Received
- October 22, 2007
- Date of Event
- October 9, 2007
- Report Date
- October 17, 2007
- Manufacturer
- HEMOSENSE, INC.
- Product Code
- JPA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
INRATIO PRECISION DATA PROVIDED BY END-USER LOT: 060452. FIRST TEST INR = 4.8. SECOND TEST INR = 4.9. MEAN = 4.9; SD = 0.07; %CV=1.46%. THE %CV IS LESS THAN OR EQUAL TO 20%. PER INTERNAL PROCEDURE, THE PRECISION PASSES THE CRITERIA FOR PRECISION. THE TEST IS CONSIDERED PRECISE AND NO FURTHER TESTING IS REQUIRED AT THIS TIME.
Description of Event or Problem · 1
CALLER ALLEGES IMPRECISION WITH INRATIO. RESULTS AS FOLLOWS: FIRST TEST INR = 4.8 SECOND TEST INR = 4.9.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | INRATIO | PROTHROMBIN TIME TEST | JPA | HEMOSENSE, INC. | 060452 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | YR |