FDA Adverse Event
Malfunction
Summary report: N
INRATIO
MDR report key: 933448
·
Received October 22, 2007
Report
- Report Number
- 2954730-2007-00504
- Event Type
- Malfunction
- Date Received
- October 22, 2007
- Date of Event
- October 2, 2007
- Report Date
- October 16, 2007
- Manufacturer
- HEMOSENSE, INC.
- Product Code
- JPA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
INRATIO PRECISION DATA PROVIDED BY END-USER LOT: 070209. FIRST TEST INR = 5.3. SECOND TEST INR = 6.1. MEAN = 5.7; SD = 0.57; %CV = 9.92%. THE %CV IS LESS THAN OR EQUAL TO 20%. PER INTERNAL PROCEDURE, THE PRECISION PASSES THE CRITERIA FOR PRECISION. THE TEST IS CONSIDERED PRECISE AND NO FURTHER TESTING IS REQUIRED AT THIS TIME.
Description of Event or Problem · 1
CALLER ALLEGES IMPRECISION WITH INRATIO. RESULTS AS FOLLOWS: FIRST TEST INR = 5.3. SECOND TEST INR = 6.1.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | INRATIO | PROTHROMBIN TIME TEST | JPA | HEMOSENSE, INC. | 070209 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | YR |