ARCHITECT TOTAL B-HCG
Report
- Report Number
- 3005094123-2019-00320
- Event Type
- Malfunction
- Date Received
- November 15, 2019
- Date of Event
- November 1, 2019
- Report Date
- February 24, 2020
- Manufacturer
- ABBOTT IRELAND DIAGNOSTICS DIVISION LONGFORD
- Product Code
- DHA
- UDI-DI
- 00380740014971
- PMA / PMN Number
- K983424
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
MANUFACTURER NAME UPDATED FROM A.I.D.D LONGFORD TO ABBOTT IRELAND DIAGNOSTICS DIVISION LONGFORD THE TICKET SEARCH DETERMINED ELEVATED COMPLAINT ACTIVITY FOR THE LOT. COMPLAINT TRENDING REPORT REVIEW DETERMINED THAT THERE IS NO ADVERSE TREND FOR THE PRODUCT. A REVIEW OF THE PRODUCT LABELING CONCLUDED THAT THE ISSUE IS SUFFICIENTLY ADDRESSED. NO CUSTOMER RETURNS WERE AVAILABLE FOR EVALUATION. A TEST PROTOCOL WAS EXECUTED USING RETAINED SAMPLES OF ARCHITECT TOTAL B-HCG REAGENT LOT 03333UI00. ALL VALIDITY AND ACCEPTANCE CRITERIA WERE MET, DEMONSTRATING THAT THE LOT IS PERFORMING ACCEPTABLY. A REVIEW OF THE PRODUCT QUALITY HISTORY FOR THE LOT NUMBER USING SEARCH OF THE CORRECTIVE AND PREVENTATIVE ACTIONS SYSTEM DID NOT IDENTIFY ISSUES ASSOCIATED WITH THE CUSTOMERS OBSERVATION. NO PRODUCT DEFICIENCY WAS IDENTIFIED.
PATIENT IDENTIFIER DOES NOT CONTAIN ENOUGH SPACES. THE ENTIRE ID IS (B)(6). AN EVALUATION IS IN PROCESS. A FOLLOWUP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE.
THE ACCOUNT GENERATED A FALSE POSITIVE ARCHITECT TOTAL BHCG OF 603.56 MIU/ML ON PATIENT SID (B)(6) THAT REPEATED NEGATIVE. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED. NO SPECIFIC PATIENT INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1121856 | ARCHITECT TOTAL B-HCG | B-HCG | DHA | ABBOTT IRELAND DIAGNOSTICS DIVISION LONGFORD | 7K78-30 | 03333UI00 | 00380740014971 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | ARCHITECT I2000SR LIST 03M74-02| ARCHITECT I2000SR LIST 03M74-02| SERIAL (B)(6)| SERIAL (B)(6) |