FDA Adverse Event Malfunction Summary report: N

BENT HOHMANN RETRACTOR

MDR report key: 9327693 · Received November 15, 2019

Report

Report Number
3005985723-2019-00806
Event Type
Malfunction
Date Received
November 15, 2019
Date of Event
October 25, 2019
Report Date
December 30, 2019
Manufacturer
MAKO SURGICAL CORP.
Product Code
OLO
UDI-DI
00848486031909
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

REPORTED EVENT: BENT AND GOUGED HOMAN CASE TYPE: NO ASSOCIATED PROCEDURE. PRODUCT EVALUATION AND RESULTS: VISUAL INSPECTION: CONFIRMS THE MAKO RETRACTOR IS BENT AND GOUGED. PRODUCT HISTORY REVIEW: REVIEW OF THE PRODUCT HISTORY RECORDS INDICATE 100 DEVICES WERE MANUFACTURED UNDER LOT NO 43245, AND ACCEPTED INTO FINAL STOCK ON 09/18/2018. NO NON-CONFORMANCES WERE IDENTIFIED DURING INSPECTION. COMPLAINT HISTORY REVIEW: A REVIEW OF COMPLAINTS IN CATSWEB AND TRACKWISE RELATED TO P/N 210191, LOT NUMBER 43245, SHOWS 00 ADDITIONAL COMPLAINTS RELATED TO THE FAILURE IN THIS INVESTIGATION. CONCLUSIONS: THE EVENT WAS CONFIRMED. PER D03391, PREVENTIVE MAINTENANCE IS WHERE AN ACTION OCCURS THAT IDENTIFIES DEVICE DETERIORATION WHICH MAY COMPROMISE FUNCTION. UNDER PM CONDITIONS NO PATIENT WAS INVOLVED AND NO ACTUAL OR POTENTIAL PATIENT HARM EXISTED FOR THE ALLEGED EVENT. NO ADDITIONAL INVESTIGATION OR SPECIFIC ACTIONS ARE REQUIRED. CORRECTIVE ACTION/PREVENTIVE ACTION: A REVIEW OF STRYKER¿S NC/CAPA DATABASE INDICATED THERE HAVE BEEN NO NC AND CAPA ASSOCIATED WITH THE PRODUCT AND FAILURE MODE REPORTED IN THIS EVENT.

Description of Event or Problem · 0

BENT AND GOUGED HOMAN. CASE TYPE: NO ASSOCIATED PROCEDURE.

Additional Manufacturer Narrative · 1

AS PART OF THE NORMAL COMPLAINT FOLLOW-UP, AN EVALUATION OF THE EVENT HAS BEEN INITIATED BY MAKO SURGICAL. A SUPPLEMENTAL REPORT WILL BE SUBMITTED WHEN ADDITIONAL INFORMATION BECOMES AVAILABLE.

Description of Event or Problem · 1

BENT AND GOUGED HOMAN. CASE TYPE: NO ASSOCIATED PROCEDURE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1124614 BENT HOHMANN RETRACTOR STEREOTAXIC DEVICE, ROBOTICS OLO MAKO SURGICAL CORP. 43245 00848486031909

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization