FDA Adverse Event Malfunction Summary report: N

BD VACUTAINER® SST¿ BLOOD COLLECTION TUBES

MDR report key: 9324493 · Received November 14, 2019

Report

Report Number
1024879-2019-01983
Event Type
Malfunction
Date Received
November 14, 2019
Date of Event
October 29, 2019
Report Date
December 17, 2019
Manufacturer
BECTON, DICKINSON & CO., (BD)
Product Code
JKA
UDI-DI
50382903679862
PMA / PMN Number
BK050036
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION: BD HAD NOT RECEIVED SAMPLES OR PHOTOS FROM THE CUSTOMER FACILITY FOR EVALUATION; THEREFORE, THE INVESTIGATION WAS LIMITED. A REVIEW OF THE DEVICE HISTORY RECORD WAS COMPLETED FOR THE INCIDENT LOT AND, BASED ON THIS REVIEW, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET; THERE WERE NO RELATED QUALITY NON-CONFORMANCES DURING MANUFACTURING OF THE PRODUCT. ERRONEOUS POTASSIUM RESULTS CAN BE CAUSED BY MANY SOURCES, INCLUDING CERTAIN PATIENT PATHOLOGICAL CONDITIONS, IMPROPER SPECIMEN COLLECTION, SPECIMEN PROCESSING, AND SPECIMEN TRANSPORT. SPECIMEN COLLECTION FACTORS THAT CAN CONTRIBUTE TO ERRONEOUS POTASSIUM RANGE FROM PROLONGED TOURNIQUET TIME AND IMPROPER VENIPUNCTURE TECHNIQUE TO TRANSFERRING A SAMPLE FROM A SYRINGE DRAW OR IV CATHETER. SPECIMEN PROCESSING FACTORS INCLUDE VIGOROUS MIXING OR SHAKING OF THE SPECIMEN, NOT ALLOWING THE SPECIMEN TO CLOT FOR THE RECOMMENDED AMOUNT OF TIME, PROLONGED CONTACT OF SERUM OR PLASMA WITH CELLS, AND EXPOSURE TO EXCESSIVE HEAT OR COLD. FINALLY, MECHANICAL TRAUMA AND OTHER ADVERSE CONDITIONS DURING TRANSPORT OF TUBES CAN RESULT IN ERRONEOUS POTASSIUM RESULTS.

Description of Event or Problem · 0

IT WAS REPORTED THAT AN UNSPECIFIED NUMBER OF BD VACUTAINER® SST¿ BLOOD COLLECTION TUBES EXPERIENCED ERRONEOUS RESULTS DURING USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: MATERIAL NO. 367986, BATCH NO. 9172629. THERE HAVE BEEN SEVERAL RESULTS WITH HIGH POTASSIUM, DATE OF EVENT: CURRENTLY UNAVAILABLE. PATIENT IDENTIFIERS: CURRENTLY UNAVAILABLE. OCCURRENCE QTY: CURRENTLY UNAVAILABLE.

Additional Manufacturer Narrative · 1

A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT AN UNSPECIFIED NUMBER OF BD VACUTAINER® SST¿ BLOOD COLLECTION TUBES EXPERIENCED ERRONEOUS RESULTS DURING USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: MATERIAL NO. 367986, BATCH NO. 9172629. THERE HAVE BEEN SEVERAL RESULTS WITH HIGH POTASSIUM, DATE OF EVENT: CURRENTLY UNAVAILABLE. PATIENT IDENTIFIERS: CURRENTLY UNAVAILABLE. OCCURRENCE QTY: CURRENTLY UNAVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1114448 BD VACUTAINER® SST¿ BLOOD COLLECTION TUBES BLOOD SPECIMEN COLLECTION DEVICE JKA BECTON, DICKINSON & CO., (BD) 9172629 50382903679862

Patients

Seq Age Sex Outcome Treatment
1 Other