FDA Adverse Event Malfunction Summary report: N

TRUE METRIX

MDR report key: 9317997 · Received November 13, 2019

Report

Report Number
1000113657-2019-10151
Event Type
Malfunction
Date Received
November 13, 2019
Date of Event
October 17, 2019
Report Date
November 13, 2019
Manufacturer
TRIVIDIA HEALTH INC
Product Code
NBW
UDI-DI
00021292007836
PMA / PMN Number
K140100
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

METER WAS RETURNED FOR EVALUATION. NO DEFECT WAS DETECTED. TEST STRIPS WERE NOT RETURNED FOR EVALUATION. MOST LIKELY UNDERLYING ROOT CAUSE: MLC-58: USER HAD AN INACCURATE REFERENCE: SELF: THE PERSON IS USING THEMSELVES AS THE REFERENCE OR HOW THEY FEEL AT THE TIME THEY RUN THE BLOOD TEST. NOTE: MANUFACTURER CONTACTED CUSTOMER IN A FOLLOW-UP CALL TO ENSURE THE REPLACEMENT PRODUCTS RESOLVED THE INITIAL CONCERN - ABLE TO ESTABLISH CONTACT WITH CUSTOMER AND STATED REPLACEMENT PRODUCT RESOLVED INITIAL CONCERN.

Description of Event or Problem · 1

CONSUMER REPORTED COMPLAINT FOR HIGH BLOOD GLUCOSE TEST RESULTS. DAUGHTER IS CALLING ON BEHALF OF THE CUSTOMER. THE CUSTOMER IS CONCERNED WITH TEST RESULTS OBTAINED OF 218, 181, 197, 157 AND 184 MG/DL. THE CUSTOMER¿S EXPECTED FASTING BLOOD GLUCOSE TEST RESULT RANGE IS 105 - 110 MG/DL. THE CUSTOMER FEELS WELL AND DID NOT REPORT ANY SYMPTOMS. MEDICAL ATTENTION IS NOT REPORTED AS A RESULT OF THE ACTUAL BLOOD GLUCOSE RESULTS. THE CUSTOMER DECLINED TO PERFORM A BACK TO BACK BLOOD TEST DURING THE CALL. THE PRODUCT IS STORED ACCORDING TO SPECIFICATION IN THE DINING ROOM. THE TEST STRIP LOT MANUFACTURER¿S EXPIRATION DATE IS 11/21/2020 AND TEST STRIPS WERE OPENED SEPTEMBER 2019. THE METER MEMORY WAS REVIEWED FOR PREVIOUS TEST RESULT HISTORY: (B)(6).