FDA Adverse Event Injury Summary report: N

COMP RVSR SHLDR GLNSP +3 36MM

MDR report key: 9316558 · Received November 13, 2019

Report

Report Number
0001825034-2019-04940
Event Type
Injury
Date Received
November 13, 2019
Date of Event
May 21, 2019
Report Date
November 13, 2019
Manufacturer
ZIMMER BIOMET, INC.
Product Code
KWS
UDI-DI
00880304475778
PMA / PMN Number
K080642
Removal / Correction Number
N/A
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). CONCOMITANT MEDICAL PRODUCTS: ITEM# 010000589; COMP RVRS 25MM BSPLT HA+ADPTR; LOT# 296770, ITEM# 115395; COMP RVS CNTRL 6.5X25MM ST/RST; LOT# 522910, ITEM# 180553; COMP LK SCR 3.5HEX 4.75X30 ST; LOT# 602720, ITEM# 180553; COMP LK SCR 3.5HEX 4.75X30 ST; LOT# 042760, ITEM# 180554; COMP LK SCR 3.5HEX 4.75X35 ST; LOT# 878350, ITEM# 113613; COMP PRIMARY STEM 13MM MICRO; LOT# 904750, ITEM# 115378; COMP RVS TRAY +10MM CO 44MM; LOT# 800040, ITEM# XL-115364; ARCOM XL 44-36 STD +3 HMRL BRG; LOT# 247330. REPORT SOURCE: (B)(6). REPORTED EVENT WAS UNABLE TO BE CONFIRMED DUE TO LIMITED INFORMATION RECEIVED FROM THE CUSTOMER. DEVICE HISTORY RECORD WAS REVIEWED AND NO DISCREPANCIES WERE FOUND. ROOT CAUSE WAS UNABLE TO BE DETERMINED AS THE NECESSARY INFORMATION TO ADEQUATELY INVESTIGATE THE REPORTED EVENT WAS NOT PROVIDED. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS. MULTIPLE MDR REPORTS WERE FILED FOR THIS EVENT, PLEASE SEE ASSOCIATED REPORTS: 0001825034-2019-02449, 0001825034-2019-04938, 0001825034-2019-04941, 0001825034-2019-04942, 0001825034-2019-04943, 0001825034-2019-04944, 0001825034-2019-04945, 0001825034-2019-04946.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT UNDERWENT A REVISION PROCEDURE APPROXIMATELY EIGHT MONTHS POST INITIAL DUE TO INFECTION. A SPACER MOLD WAS TEMPORARILY IMPLANTED AND REMOVED AT A LATER DATE. NO FURTHER INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1108792 COMP RVSR SHLDR GLNSP +3 36MM PROSTHESIS, EXTREMITIES KWS ZIMMER BIOMET, INC. N/A 020060 00880304475778

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization| R