FDA Adverse Event Malfunction Summary report: N

SEDECAL

MDR report key: 931387 · Received October 22, 2007

Report

Report Number
9617251-2007-00002
Event Type
Malfunction
Date Received
October 22, 2007
Date of Event
September 21, 2007
Report Date
October 22, 2007
Manufacturer
SEDECAL S.A.
Product Code
IZO
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE DIAGNOSTIC X-RAY SYSTEM EXPERIENCED UNCOMMANDED MOVEMENT OF THE TUBE-HEAD/BUCKY ASSEMBLY. THE ACTUAL CAUSE OF THIS MOVEMENT IS UNKNOWN AS OF THE DATE OF THIS REPORT. HOWEVER AND UPGRADE TO THE CONTROL LOGIC HAS BEEN MADE TO SYSTEMS MANUFACTURED DURING AND AFTER APRIL 2007 HAS BEEN IMPLEMENTED. THIS UPGRADE PREVENTS UNCOMMANDED SYSTEM MOVEMENT. THE SYSTEM THAT MALFUNCTIONS HAD NOT RECEIVED THIS UPGRADE. THE SYSTEM WAS REPLACED AND THE SYSTEM THAT MALFUNCTIONED HAS BEEN SENT TO SEDECAL USA FOR FURTHER EVALUATION.

Description of Event or Problem · 1

THE POSITIONER WAS IN THE UNDERTABLE POSITION, SID AT 100CM AND WITH A PATIENT ON A QUANTUM TABLE. WHILE THE X-RAY TECHNICIAN WAS POSITIONING THE PATIENT ON THE TABLE, THEY NOTICED THE U-ARM POSITIONER START TO MOVE IN THE DOWNWARD POSITION BY ITSELF. THE TECHNICIAN QUICKLY MOVE THE PATIENT OFF THE TABLE TO AVOID THE PATIENT BEING HIT BY THE MOVING TUBE/COLLIMATOR ASSEMBLY. THEY THEN NOTICED THE POSITIONER START TO TRAVEL IN THE UPWARD POSITION ON ITS OWN, WITH THE TABLE STILL OVER THE BUCKY. IT ACTUALLY LIFTED THE SEVERAL HUNDRED POUND TABLE COMPLETELY OFF THE FLOOR. THE POSITIONER CONTINUED TO MOVE UP PAST THE HEIGHT MOTOR LIMIT SWITCHES, CRUSHING THE MECHANICAL STOPS. THE CARRIAGE CABLE CAME COMPLETELY OFF THE PULLEY, WRAPPING AROUND THE HEIGHT POTENTIOMETER BRACKET. BESIDES THE MAJOR SAFETY ISSUE HERE, SEVERE DAMAGE OCCURRED TO THE POSITIONER. THE PATIENT WAS NOT INJURED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SEDECAL DIAGNOSTIC X-RAY SYSTEM IZO SEDECAL S.A. SEDECAL X-PLUS

Patients

Seq Age Sex Outcome Treatment
1 YR