CUSA EXCEL 36KHZ STRAIGHT HANDPIECE
Report
- Report Number
- 3006697299-2019-00126
- Event Type
- Malfunction
- Date Received
- November 12, 2019
- Report Date
- October 15, 2019
- Manufacturer
- INTEGRA LIFESCIENCES(IRELAND)
- Product Code
- LFL
- PMA / PMN Number
- K141674
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
THE DEVICE WAS RETURNED FOR EVALUATION. THE DHR DOCUMENTATION WAS REVIEWED AND NO ANOMALIES THAT COULD BE ASSOCIATED WITH THE COMPLAINT INCIDENT WAS OBSERVED. A SERIAL NUMBER REVIEW FOR THE HANDPIECE CONFIRMED THIS IS A SINGLE COMPLAINT FOR THE AFFECTED DEVICE. THE COMPLAINT WAS VERIFIED AS VALID. THE TECHNICIAN FOUND THE HANDPIECE WAS OUT OF SPECIFICATION BETWEEN 05NOV2019 AND 30JUN2020, APPROXIMATELY 2,200 MDRS SUBMITTED ELECTRONICALLY BY INTEGRA LIFESCIENCES VIA TRACKWISE, INTEGRA'S COMPLAINT HANDLING SYSTEM, WERE NOT RECEIVED BY CDRH DUE TO A COMPUTER SYSTEM ISSUE. WITHIN THIS TIME PERIOD, AN ERROR WITH INTEGRA'S MIDDLEWARE, WHICH FACILITATES COMMUNICATIONS BETWEEN TRACKWISE AND THE FDA SYSTEM, CAUSED THE COMPLAINT RECORDS TO CLOSE AND INDICATE WE HAD RECEIVED AN ACKNOWLEDGEMENT 3 FROM THE FDA WHEN WE HAD NOT. INTEGRA INTERPRETED THE ACKNOWLEDGEMENT AS A SUCCESSFUL SUBMISSION; HOWEVER, SUBSEQUENT INVESTIGATION REVEALED THE ACKNOWLEDGEMENT 3 RECEIVED WAS FROM OUR MIDDLEWARE AND NOT FROM THE FDA (THESE ACKNOWLEDGEMENTS HAVE BEEN RETAINED AS PART OF THE DOCUMENTATION OF THE MDR). THE MALFUNCTION WAS RELATED TO THE RELOCATION OF THE TRACKWISE APPLICATION TO A NEW DATA CENTER DURING THE TRANSITION OF INTEGRA'S CORPORATE HEADQUARTERS FROM PLAINSBORO, NJ TO PRINCETON, NJ. PREVIOUSLY, INTEGRA HAD BEEN SUCCESSFULLY RECEIVING ACKNOWLEDGEMENTS 1, 2, AND 3 FROM THE FDA, AND OUR RECORDS REFLECT THESE ACKNOWLEDGEMENTS, INCLUDING THE DATE AND TIME STAMPS. CAPA PR 229048 AND NC 20-011 HAVE BEEN OPENED BY INTEGRA TO FURTHER INVESTIGATE THE NONCONFORMANCE AND DEVELOP A CORRECTIVE ACTION PLAN. THE MIDDLEWARE ERROR HAS BEEN CORRECTED, AND INTEGRA HAS FILED MDRS SINCE THE CORRECTION AND VERIFIED THAT THE APPROPRIATE ACKNOWLEDGEMENTS HAVE BEEN RECEIVED FROM THE FDA. INTEGRA IS RESUBMITTING ALL IMPACTED MDR REPORTS FOR THE TIME PERIOD 05NOV2019 THROUGH 30JUN2020. INTEGRA LIFESCIENCES CONTACTED (B)(4), DIRECTOR OF REGULATORY PROGRAMS, OFFICE OF PRODUCT EVALUATION AND QUALITY AND (B)(4), ASSISTANT DIRECTOR, MDR TEAM, OFFICE OF PRODUCT EVALUATION AND QUALITY ON JULY 8-9, 2020 TO REPORT THESE ISSUES REGARDING MDR REPORTS.
N/A.
THE DEVICE WAS NOT YET RETURNED TO THE MANUFACTURER FOR ANALYSIS. THE PLANT INVESTIGATION IS IN PROGRESS AND A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED UPON COMPLETION OF THE INVESTIGATION. LINKED TO MFG REPORT NO.: 3006697299-2019-00127.
THIS IS 1 OF 2 REPORTS. A SALES REPRESENTATIVE REPORTED THAT THE C2600 CUSA EXCEL 23KHZ STRAIGHT HANDPIECE FAILED DURING TESTING. THERE WAS NO KNOWN PATIENT INVOLVEMENT OR DELAY IN SURGERY. AN ADDITIONAL INFORMATION RECEIVED ON 21OCT2019 AND 28OCT2019 INDICATING THAT THE HANDPIECES WERE BEING USED DURING DEMONSTRATION AND NOT DURING PATIENT CARE. THEY WERE FOUND TO BE BROKEN AND OVER TORQUED WITH TWISTED TRANSDUCERS WHEN A FIELD SERVICE TECHNICIAN WAS CHECKING THE CUSA CONSOLE FOR PREVENTATIVE MAINTENANCE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1106316 | CUSA EXCEL 36KHZ STRAIGHT HANDPIECE | ULTRASONIC SURGICAL PRODUCTS | LFL | INTEGRA LIFESCIENCES(IRELAND) |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |