Access hsTnI
Report
- Report Number
- 2122870-2019-01139
- Event Type
- Injury
- Date Received
- November 12, 2019
- Date of Event
- October 31, 2019
- Report Date
- November 12, 2019
- Manufacturer
- BECKMAN COULTER
- Product Code
- MMI
- UDI-DI
- 15099590693183
- PMA / PMN Number
- K121790
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AS
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
THE DATA SUBMITTED BY THE CUSTOMER TO BECKMAN COULTER FOR INVESTIGATION SUPPORTS EVIDENCE OF REPEATABLE ERRONEOUSLY ELEVATED HSTNI PATIENT RESULT OBTAINED ON ACCESS2 ANALYZER S/N (B)(4): THE ORIGINAL RESULT WAS 87 NG/L ON (B)(6) 2019 AND DUPLICATE RESULTS OF UNCENTRIFUGED SAMPLE WERE RESPECTIVELY 3.3 AND 3.5 NG/L ON (B)(6) 2019. THE SAMPLE WAS COLLECTED IN A RAPID SERUM TUBE LOT 190209, IT WAS FULL AND CLEAR. CALIBRATION PASSED ON (B)(6) 2019 WITH REAGENT LOT 832524 AND CALIBRATOR LOT 832156. SYSTEM CHECK PASSED ON (B)(6) 2019. QC WERE PASSING WITHIN THE LABORATORY¿S ESTABLISHED RANGES. NO HARDWARE ERRORS OR OTHER ASSAY ISSUES WERE REPORTED IN CONJUNCTION WITH THIS EVENT. THE CAUSE OF THIS EVENT IS UNKNOWN. (B)(4).
THE CUSTOMER REPORTED OBTAINING ERRONEOUS ELEVATED TROPONIN I (ACCESS HSTNI) PATIENT RESULTS INVOLVING THE LABORATORY'S ACCESS 2 IMMUNOASSAY SYSTEM ACCESS CLINICAL SYSTEM (SERIAL NUMBER (B)(4)). RESULTS WERE DISCORDANT WITH THE ABBOTT I-STAT CTNI METHOD AND THE PATIENT CLINICAL FILE. CALIBRATION, QC (QUALITY CONTROL) AND SYSTEM CHECK WERE WITHIN SYSTEM AND ASSAY SPECIFICATIONS. ADDITIONAL INFORMATION ON THE PATIENT¿S DEMOGRAPHICS SUCH AS AGE OR DATE OF BIRTH WAS NOT SUPPLIED. INVESTIGATIONS REQUIRED. THERE WAS A CHANGE IN PATIENT TREATMENT BASED ON THE ORIGINAL ERRONEOUSLY ELEVATED HSTNI RESULT. PATIENT TREATED FROM (B)(6) 2019 WITH THE FOLLOWING MEDICATION BASED ON INITIAL TROPONIN RESULT AT 15:50 (B)(6) 2019: CLOPIDOGREL 300 MG LOADING DOSE + 3 DOSES OF 75 MG. ASPIRIN 300 MG LOADING DOSE + 3 DOSES OF 100 MG.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1104421 | Access hsTnI | IMMUNOASSAY METHOD, TROPONIN SUBUNIT | MMI | BECKMAN COULTER | ACCESS HSTNI 2X50 DET | 832524 | 15099590693183 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |