FDA Adverse Event Injury Summary report: N

STEM: AMISTEM H HA COATED STD STEM SIZE 1

MDR report key: 9308985 · Received November 12, 2019

Report

Report Number
3005180920-2019-00964
Event Type
Injury
Date Received
November 12, 2019
Date of Event
October 22, 2019
Report Date
November 12, 2019
Manufacturer
MEDACTA INTERNATIONAL SA
Product Code
LZO
UDI-DI
07630030804076
PMA / PMN Number
K093944
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 1

BATCH REVIEW PERFORMED ON 29 OCTOBER 2019: LOT 1903333: 153 ITEMS MANUFACTURED AND RELEASED ON 26-JUN-2019. EXPIRATION DATE: 2024-06-17. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, 55 ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY SIMILAR REPORTED EVENT. CLINICAL EVALUATION: FEW WEEKS AFTER CEMENTLESS TOTAL HIP ARTHROPLASTY, THE PATIENT REPORTED PAIN DUE TO A FEMORAL FRACTURE. DURING REHABILITATION PERIOD, A SUDDEN MOVEMENT ON A WEAKENED BONE DUE TO NORMAL FEMORAL PREPARATION MAY FAVOUR THE OCCURRENCE OF EARLY FRACTURE. THERE IS NO REASON TO SUSPECT A FAULTY DEVICE.

Description of Event or Problem · 1

22 DAYS AFTER PRIMARY SURGERY THE SURGEON REVISED THE PATIENT HIP FOR PERI-PROSTHETIC FEMUR FRACTURE. THE SURGEON REPLACED SUCCESSFULLY THE STEM USING A COMPETITOR DEVICE AND FIXED THE FRACTURE WITH WIRES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1104567 STEM: AMISTEM H HA COATED STD STEM SIZE 1 HIP CEMENTLESS STEM LZO MEDACTA INTERNATIONAL SA 1903333 07630030804076

Patients

Seq Age Sex Outcome Treatment
1 71 YR Required Intervention