FDA Adverse Event Malfunction Summary report: N

SYRINGE S2 5ML 22GA 1-1/4IN

MDR report key: 9306062 · Received November 11, 2019

Report

Report Number
3002682307-2019-00603
Event Type
Malfunction
Date Received
November 11, 2019
Date of Event
October 25, 2019
Report Date
December 12, 2019
Manufacturer
BECTON DICKINSON, S.A.
Product Code
FMF
PMA / PMN Number
N/A
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION SUMMARY: BD HAS BEEN PROVIDED WITH PHOTO FOR CATALOG 301942 LOT 1901146 TO INVESTIGATE FOR THIS RECORD. VISUAL EXAMINATION OF THE PHOTO SHOWS THE LABEL WITH NO LOT NUMBER ON THE PRINTED LABEL. AS A RESULT, BD WAS ABLE TO VERIFY THE REPORTED ISSUE. BD CONCLUDED THAT THE CAUSE OF THE PROBLEM WAS RELATED WITH SOME UNEXPECTED PROBLEM IN THE PRINTING MACHINE. AS A CONSEQUENCE, ONLY THE MIDDLE OF THE REPORTED SHELF CARTON LABEL WAS PRINTED. CONSIDERING OUR IN-COMING AND IN-PROCESS INSPECTION AND SINCE THIS IS THE FIRST TIME THIS LOT IS REPORTED FOR THIS DEFECT, NO CORRECTIVE ACTIONS ARE REQUIRED AT THIS TIME. A REVIEW OF THE DEVICE HISTORY RECORD REVEALED NO IRREGULARITIES DURING THE MANUFACTURE OF THE REPORTED LOT.

Description of Event or Problem · 0

IT WAS REPORTED THAT LABEL ERROR OCCURRED WITH A SYRINGE S2 5ML 22GA 1-1/4IN. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, "NO LOT NUMBER ON THE MIDDLE PACKAGE." 100 OCCURRENCES REPORTED.

Additional Manufacturer Narrative · 1

A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT LABEL ERROR OCCURRED WITH A SYRINGE S2 5ML 22GA 1-1/4IN. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, "NO LOT NUMBER ON THE MIDDLE PACKAGE." 100 OCCURRENCES REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1097484 SYRINGE S2 5ML 22GA 1-1/4IN SYRINGE FMF BECTON DICKINSON, S.A. 1901146

Patients

Seq Age Sex Outcome Treatment
1 Other