FDA Adverse Event Injury Summary report: N

EXACT COUCH

MDR report key: 930043 · Received October 19, 2007

Report

Report Number
2914292-2007-00039
Event Type
Injury
Date Received
October 19, 2007
Date of Event
October 8, 2007
Report Date
October 8, 2007
Manufacturer
VARIAN MEDICAL SYSTEMS
Product Code
JAI
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THIS INCIDENT IS RECOGNIZED AS USE ERROR BY VARIAN MEDICAL SYSTEMS. LABELING PLACED ON THE COUCH, AS WELL AS IN THE USER MANUAL, PROVIDE INSTRUCTION AND HAZARD WARNINGS (WITH RESPECT TO HAND PLACEMENT), WHICH HELP TO ENSURE THE SAFE OPERATION OF THE EQUIPMENT. HOWEVER, CAPA HAS BEEN OPENED TO INVESTIGATE POSSIBLE PRODUCT IMPROVEMENT TO FURTHER MITIGATE THIS DOCUMENTED RISK.

Description of Event or Problem · 1

OPERATOR A WAS HOLDING THE COUCH AT THE BACK OF COUCH TOP- FINGERS IN-BETWEEN THE TABLETOP EDGE AND MOVABLE CARBON FIBER RAIL. OPERATOR B WAS ASSISTING IN MOVING THE COUCH TOP WITH PT LOADED IN THE LONGITUDINAL DIRECTION. WITH OPERATOR A'S FIRST FINGERS IN THE LOCATION STATED ABOVE, A PINCH POINT EXISTED BETWEEN THE MOVING CARRIAGE AND THE STATIONARY LATERAL BASE. THE COUCH TOP WAS MOVED IN TOWARDS THE STAND TOO FAST AND BROKE THE BONE ON THE FINGER OF THERAPIST A.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 EXACT COUCH PATIENT POSITIONING TABLE JAI VARIAN MEDICAL SYSTEMS EXACT COUCH NA