FDA Adverse Event Injury Summary report: N

MENTOR SMOOTH ROUND MODERATE PLUS PROFILE

MDR report key: 9293605 · Received November 7, 2019

Report

Report Number
1645337-2019-23381
Event Type
Injury
Date Received
November 7, 2019
Date of Event
October 9, 2019
Report Date
October 10, 2019
Manufacturer
MENTOR TEXAS
Product Code
FWM
UDI-DI
00081317001423
PMA / PMN Number
P990075
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

ON 12/5/2019, THE PRODUCT INVESTIGATION WAS COMPLETED. DEVICE INVESTIGATION SUMMARY: DURING VISUAL EVALUATION OF THE SAMPLE IT WAS OBSERVED A TEAR LOCATED IN THE UNION BETWEEN VALVE AND SHELL. MICROSCOPIC EXAMINATION WAS PERFORMED. NO EVIDENCE OF INSTRUMENT DAMAGE WAS OBSERVED. NO OTHER ANOMALIES WERE DISCOVERED. THE SECOND PRODUCT RECEIVED IS RELATED TO A CONCOMITANT DEVICE, THERE ARE NEITHER DEFICIENCIES ALLEGED NOR PATIENT INJURIES. THEREFORE NO FURTHER INVESTIGATION IS REQUIRED. DEFLATION IS A KNOWN COMPLICATION ASSOCIATED WITH MENTOR MAMMARY PROSTHESES. POSSIBLE CAUSES OF DEFLATION OF SALINE-FILLED IMPLANTS INCLUDE, BUT ARE NOT LIMITED TO THE FOLLOWING EVENTS: EXCESSIVE STRESSES OR MANIPULATIONS AS MAY OCCUR DURING NORMAL, DAILY ROUTINES. THE MANUFACTURING RECORDS WERE REVIEWED, AND THE MANUFACTURING CRITERIA WERE MET PRIOR TO THE RELEASE OF THIS LOT. IT SHOULD BE NOTED THAT AS PART OF OUR QUALITY PROCESS, EACH DEVICE IS VISUALLY INSPECTED DURING MANUFACTURING TO ENSURE THE DEVICE MEETS THE REQUIRED SPECIFICATIONS PRIOR TO SHIPMENT. MANUFACTURER REFERENCE NUMBER: (B)(4).

Additional Manufacturer Narrative · 1

THE MENTOR FAILURE ANALYSIS LAB HAS RECEIVED THE DEVICE FOR EVALUATION. THE ANALYSIS HAS BEGUN BUT IS NOT COMPLETE AT THIS TIME. WHEN THE INVESTIGATIONAL ANALYSIS HAS BEEN COMPLETED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. A MANUFACTURING RECORD EVALUATION (MRE) WAS PERFORMED, AND NO ANOMALIES WERE FOUND RELATED TO THIS COMPLAINT. IN ADDITION, THE MRE VERIFIES THAT THE DEVICE WAS MANUFACTURED IN ACCORDANCE WITH DOCUMENTED SPECIFICATION AND PROCEDURES. REASON FOR DEVICE EXPLANT AND/OR REOPERATION: DEFLATION. CONCOMITANT PRODUCTS: (LEFT) 375CC MENTOR SMOOTH ROUND MODERATE PLUS PROFILE SALINE CATALOG: 3502375 LOT: 5983557 SN: (B)(4). MANUFACTURER¿S REFERENCE NUMBER: (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT A (B)(6) YEAR OLD FEMALE PATIENT WHO UNDERWENT PRIMARY BREAST AUGMENTATION WITH TWO 375CC MENTOR SMOOTH ROUND MODERATE PLUS PROFILE SALINE BREAST PROSTHESES, SUFFERED RIGHT SIDE DEFLATION, POST-OPERATIVELY. AS A RESULT, THE PATIENT UNDERWENT IMPLANT EXPLANTATION ON (B)(6) 2019.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1088315 MENTOR SMOOTH ROUND MODERATE PLUS PROFILE PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE FWM MENTOR TEXAS 5983557 00081317001423

Patients

Seq Age Sex Outcome Treatment
1 50 YR Required Intervention