FDA Adverse Event Malfunction Summary report: N

CNS-6801A

MDR report key: 9293523 · Received November 7, 2019

Report

Report Number
8030229-2019-00629
Event Type
Malfunction
Date Received
November 7, 2019
Date of Event
October 15, 2019
Report Date
March 17, 2020
Manufacturer
NIHON KOHDEN CORPORATION
Product Code
MHX
UDI-DI
04931921131640
PMA / PMN Number
K102376
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 0

DETAILS OF COMPLAINT: ON (B)(6) 2019 CUSTOMER STATED THAT COMMUNICATION LOSS WAS REPORTED BETWEEN RNS DEVICES AND ALL GZ TRANSMITTERS. CNS MONITORS WERE NOT AFFECTED. THE DEVICE LISTED ON THIS COMPLAINT IS THE NETWORK SERVER: A/PWREDGE-R640 (S/N (B)(6). INVESTIGATION SUMMARY: DUE TO LIMITED INFORMATION AVAILABLE, THE ROOT CAUSE OF THE REPORTED COMMUNICATION LOSS ISSUE COULD NOT BE DETERMINED. ADDITIONAL MODEL INFORMATION: D11 & C2: THE FOLLOWING DEVICES WERE USED IN CONJUNCTION WITH THE CNS AND ARE DEVICES THAT ALSO EXPERIENCED FAILURE. ATTEMPTS TO OBTAIN THE FOLLOWING INFORMATION WERE MADE, BUT NOT PROVIDED: CNS'S - MODEL: NI. S/N: NI. APPROXIMATE AGE OF THE DEVICE: NO SERIAL NUMBER WAS PROVIDED, SO THE AGE OF THE DEVICE IS UNKNOWN. DEVICE MANUFACTURER DATE: NI. UNIQUE IDENTIFIER (UDI) #: NI. RNS'S - MODEL: NI. S/N: NI. APPROXIMATE AGE OF THE DEVICE: NO SERIAL NUMBER WAS PROVIDED, SO THE AGE OF THE DEVICE IS UNKNOWN. DEVICE MANUFACTURER DATE: NI. UNIQUE IDENTIFIER (UDI) #: NI. TRANSMITTERS - MODEL: NI. S/N: NI. APPROXIMATE AGE OF THE DEVICE: NO SERIAL NUMBER WAS PROVIDED, SO THE AGE OF THE DEVICE IS UNKNOWN. DEVICE MANUFACTURER DATE: NI. UNIQUE IDENTIFIER (UDI) #: NI.

Description of Event or Problem · 0

THE BIOMEDICAL ENGINEER REPORTED THAT COMMUNICATION LOSS OCCURRED ON SITE FOR ABOUT 30-45 SECONDS AND THAT IT AFFECTED A COUPLE OF THEIR CENTRAL NURSE'S STATIONS (CNS'S), ALL OF THEIR REMOTE NETWORK STATIONS (RNS'S), AND ALL OF THEIR TRANSMITTERS.

Additional Manufacturer Narrative · 1

THE BIOMEDICAL ENGINEER REPORTED THAT COMMUNICATION LOSS OCCURRED ON SITE FOR ABOUT 30-45 SECONDS AND THAT IT AFFECTED A COUPLE OF THEIR CENTRAL NURSE'S STATIONS (CNS'S), ALL OF THEIR REMOTE NETWORK STATIONS (RNS'S), AND ALL OF THEIR TRANSMITTERS. CITS TEAM FOUND THAT THERE WAS A DISCONNECT ON BOTH .10 AND .30 SEGMENT AND THAT THERE WAS A TIME ADJUSTMENT OF TWO MINUTES THAT HAPPENED ON THE PRIMARY HOST SERVER THAT WAS BEING USED AS A TIME SYNC. THEY ALSO FOUND A COMMON POINT FOR THE HOST SERVERS AND ALL OF THE MENTIONED DEVICES LOCATED ON THE 4TH FLOOR, MANAGED BY A SINGLE CISCO SWITCH. NO PATIENT HARM OR INJURY WAS REPORTED. CITS TEAM ADVISED THE CUSTOMER TO VERIFY THE TIME SETTINGS ON ALL OF THEIR CNS'S IN ORDER FOR FUTURE TROUBLESHOOTING TO TAKE PLACE. NIHON KOHDEN CONTINUES TO INVESTIGATE THE REPORTED EVENT. NIHON KOHDEN WILL SUBMIT A SUPPLEMENTAL REPORT IN ACCORDANCE WITH 21 CFR SECTION 803.56 WHEN ADDITIONAL INFORMATION BECOMES AVAILABLE. ADDITIONAL MODEL INFORMATION: THE FOLLOWING DEVICES WERE USED IN CONJUNCTION WITH THE CNS AND ARE DEVICES THAT ALSO EXPERIENCED FAILURE. ATTEMPTS TO OBTAIN THE FOLLOWING INFORMATION WERE MADE, BUT NOT PROVIDED: CNS'S - MODEL: NI, S/N: NI. APPROXIMATE AGE OF THE DEVICE: NO SERIAL NUMBER WAS PROVIDED, SO THE AGE OF THE DEVICE IS UNKNOWN. DEVICE MANUFACTURER DATE: NI. UNIQUE IDENTIFIER (UDI) #: NI. RNS'S - MODEL: NI, S/N: NI, APPROXIMATE AGE OF THE DEVICE: NO SERIAL NUMBER WAS PROVIDED, SO THE AGE OF THE DEVICE IS UNKNOWN. DEVICE MANUFACTURER DATE: NI. UNIQUE IDENTIFIER (UDI) #: NI. TRANSMITTERS - MODEL: NI, S/N: NI, APPROXIMATE AGE OF THE DEVICE: NO SERIAL NUMBER WAS PROVIDED, SO THE AGE OF THE DEVICE IS UNKNOWN. DEVICE MANUFACTURER DATE: NI. UNIQUE IDENTIFIER (UDI) #: NI.

Description of Event or Problem · 1

THE BIOMEDICAL ENGINEER REPORTED THAT COMMUNICATION LOSS OCCURRED ON SITE FOR ABOUT 30-45 SECONDS AND THAT IT AFFECTED A COUPLE OF THEIR CENTRAL NURSE'S STATIONS (CNS'S), ALL OF THEIR REMOTE NETWORK STATIONS (RNS'S), AND ALL OF THEIR TRANSMITTERS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1085938 CNS-6801A CENTRAL MONITOR SYSTEM MHX NIHON KOHDEN CORPORATION CNS-6801A NA 04931921131640

Patients

Seq Age Sex Outcome Treatment
1 ALL OF THE RNS'S| ALL OF THE RNS'S| ALL TRANSMITTERS| ALL TRANSMITTERS| VARIETY OF CNS'S| VARIETY OF CNS'S