FDA Adverse Event Injury Summary report: N

SAPIEN VALVE UNKNOWN

MDR report key: 9282600 · Received November 5, 2019

Report

Report Number
2015691-2019-04108
Event Type
Injury
Date Received
November 5, 2019
Date of Event
September 25, 2019
Report Date
October 11, 2019
Manufacturer
EDWARDS LIFESCIENCES
Product Code
NPT
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 0

CORRECTED DATA: F10, H6. REFERENCE CAPA-20-00141.

Additional Manufacturer Narrative · 1

THE DATE OF THE EVENT IS UNKNOWN; HOWEVER, THE INFORMATION WAS PRESENTED ON SEPTEMBER 25, 2019. THEREFORE, THE PRESENTATION DATE WAS USED AS THE OCCURRENCE DATE. IT IS UNKNOWN IF A SAPIEN, SAPIEN XT OR SAPIEN 3 VALVE WAS IMPLANTED. THE PMA NUMBERS ARE P110021 FOR SAPIEN, P130009 FOR SAPIEN XT AND P100041 FOR SAPIEN 3. PER THE INSTRUCTIONS FOR USE (IFU), PARAVALVULAR LEAK (PVL) IS A POTENTIAL ADVERSE EVENT ASSOCIATED WITH BIOPROSTHETIC HEART VALVES AND THE TRANSCATHETER AORTIC VALVE REPLACEMENT (TAVR) PROCEDURE. THE PATIENT SCREENING MANUAL AND THE PROCEDURE DIDACTIC IDENTIFY SEVERAL PROCEDURAL AND ANATOMICAL FACTORS WHICH COULD CONTRIBUTE TO PVL, INCLUDING DEVICE MALPOSITION, INACCURATE MEASUREMENT OF THE NATIVE VALVE ANNULUS, UNEVEN DISTRIBUTION OF CALCIUM ON THE NATIVE VALVE, BULKY OR SEVERE CALCIFICATION, AN ELLIPTICAL ANNULUS SHAPE AND VALVE UNDER-SIZING. PHYSICIANS ARE EXTENSIVELY TRAINED BY EDWARDS BEFORE THEY ARE QUALIFIED TO USE THE SAPIEN THV. THE PATIENT SCREENING MANUAL INSTRUCTS THE OPERATOR ON PROPER AORTIC VALVE AND ROOT ASSESSMENT, INCLUDING THE USE OF ECHO, AORTOGRAM AND CT TO APPROPRIATELY MEASURE THE ANNULUS DIAMETER, CONTENT AND DISTRIBUTION OF CALCIUM, AND LEAFLET CHARACTERISTICS. CONTRAINDICATIONS, IMPORTANT CONSIDERATIONS WHEN ASSESSING THE VALVE, AND CHOOSING THE PROPER THV ARE ALSO DISCUSSED. THE THV TRAINING MANUALS ALSO INSTRUCT THE OPERATOR ON PROPER POSITIONING AND DEPLOYMENT OF THE VALVE, INCLUDING ALL PROCEDURAL AND ANATOMICAL CONSIDERATIONS. DEVICE PREPARATION, APPROACH, DEPLOYMENT, IMAGING, PROCEDURE-SPECIFIC TRAINING MANUALS AND PROCTORED PROCEDURES ARE ALSO INCLUDED. IN THIS CASE, THERE WAS NO ALLEGATION OR INDICATION A DEVICE MALFUNCTION CONTRIBUTED TO THIS ADVERSE EVENT. PER THE REPORT, ¿GROSS UNDER SIZING¿ WAS CONSIDERED TO BE THE CAUSE FOR THE PVL AND VALVE ROCKING. THE IFU AND TRAINING MANUALS HAVE BEEN REVIEWED AND NO INADEQUACIES HAVE BEEN IDENTIFIED WITH REGARDS TO WARNINGS, CONTRAINDICATIONS, AND THE DIRECTIONS/CONDITIONS FOR THE SUCCESSFUL USE OF THE DEVICE. COMPLAINT HISTORIES FOR ALL REPORTED EVENTS ARE REVIEWED AGAINST TRENDING CONTROL LIMITS ON A MONTHLY BASIS, AND ANY EXCURSIONS ABOVE THE CONTROL LIMITS ARE ASSESSED AND DOCUMENTED AS PART OF THIS MONTHLY REVIEW. NO CORRECTIVE OR PREVENTATIVE ACTIONS ARE REQUIRED.

Description of Event or Problem · 1

AS REPORTED THROUGH THE TCT 2019 CONFERENCE, A 26MM EDWARDS¿S VALVE HAD SEVERE DYNAMIC PARAVALVULAR LEAK (PVL) AND ROCKING. ¿GROSS UNDER SIZING¿ WAS CONSIDERED TO BE THE CAUSE FOR THE PVL AND VALVE ROCKING. NO ADDITIONAL INFORMATION WAS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1071351 SAPIEN VALVE UNKNOWN AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIVERED NPT EDWARDS LIFESCIENCES SAPIEN VALVE UNKNOWN

Patients

Seq Age Sex Outcome Treatment
1