THERMOCOOL® SMART TOUCH® SF BI-DIRECTIONAL NAVIGATION CATHETER

MDR report key: 9282423 · Received November 5, 2019

Report

Report Number
2029046-2019-03855
Event Type
Malfunction
Date Received
November 5, 2019
Date of Event
October 11, 2019
Report Date
October 14, 2019
Manufacturer
BIOSENSE WEBSTER INC
Product Code
LPB
UDI-DI
10846835010183
PMA / PMN Number
P030031
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

THE BIOSENSE WEBSTER, INC. PRODUCT ANALYSIS LAB RECEIVED THE DEVICE FOR EVALUATION ON (B)(6) 2019. UPON INITIAL VISUAL INSPECTION, IT WAS REPORTED THAT THERE WAS DARK RED / BLACK MATERIAL ON THE DOME OF THE CATHETER. THE ANALYSIS HAS BEGUN BUT IS NOT COMPLETED AT THIS TIME. WHEN THE INVESTIGATIONAL ANALYSIS HAS BEEN COMPLETED, A SUPPLEMENTAL 3500A REPORT WILL BE SUBMITTED. MANUFACTURER'S REFERENCE # (B)(4).

Additional Manufacturer Narrative · 0

IT WAS REPORTED THAT A PATIENT UNDERWENT AN ATRIAL FIBRILLATION (AFIB) PROCEDURE WITH A THERMOCOOL® SMART TOUCH® SF BI-DIRECTIONAL NAVIGATION CATHETER AND A THROMBUS ISSUE OCCURRED. INVESTIGATION ANALYSIS WAS COMPLETED ON (B)(6) 2019. THE DEVICE WAS VISUALLY INSPECTED AND DARK RED MATERIAL WAS FOUND ON THE DOME. DURING THE SECOND VISUAL INSPECTION, THROMBUS WAS IDENTIFIED AS THE MATERIAL ON THE DOME. AN ELECTRICAL TEST WAS PERFORMED ON THE CATHETER AND IT WAS FOUND WITHIN SPECIFICATIONS, NO ELECTRICAL MALFUNCTION WAS OBSERVED. ADDITIONALLY, THE CATHETER WAS TESTED ON THE GENERATOR AND THE TEMPERATURE AND IMPEDANCE VALUES WERE OBSERVED WITHIN SPECIFICATIONS. A COOL FLOW PUMP TEST WAS PERFORMED, AND IT WAS FOUND WITHIN SPECIFICATIONS, THE CATHETER WAS IRRIGATING CORRECTLY, NO IRRIGATION ISSUES WERE OBSERVED. A MANUFACTURING RECORD EVALUATION WAS PERFORMED, AND NO NON-CONFORMITIES RELATED TO THE COMPLAINT WAS FOUND DURING THE REVIEW. THE CUSTOMER COMPLAINT WAS CONFIRMED. THE ROOT CAUSE OF THE THROMBUS FOUND ON THE DOME CANNOT BE DETERMINED HOWEVER, THIS COULD BE RELATED TO THE PROCEDURE. ADDITIONAL INFORMATION WAS RECEIVED ON DECEMBER (B)(6)2019. SMART ABLATE GENERATOR WAS USED DURING THE PROCEDURE. THERE WERE NO ERROR MESSAGES AND THE PHYSICIAN/USER DID NOT SEE ANY PRODUCT PROBLEM. D11 HAS BEEN UPDATED TO INCLUDE THE SMARTABLATE GENERATOR. MANUFACTURER'S REF. NO: (B)(4).

Additional Manufacturer Narrative · 1

(B)(6). SINCE THE PRODUCT WAS NOT RETURNED FOR ANALYSIS, NO PRODUCT FAILURE ANALYSIS CAN BE CONDUCTED, AND NO DETERMINATION OF POSSIBLE CONTRIBUTING FACTORS COULD BE MADE. IF ADDITIONAL INFORMATION IS RECEIVED REGARDING THIS EVENT, A SUPPLEMENTAL 3500A REPORT WILL BE SUBMITTED TO THE FDA. A MANUFACTURING RECORD EVALUATION WAS PERFORMED FOR THE FINISHED DEVICE 30255929M NUMBER, AND NO INTERNAL ACTIONS RELATED TO THE COMPLAINT WAS FOUND DURING THE REVIEW. MANUFACTURER'S REFERENCE # (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT A PATIENT UNDERWENT AN ATRIAL FIBRILLATION (AFIB) PROCEDURE WITH A THERMOCOOL® SMART TOUCH® SF BI-DIRECTIONAL NAVIGATION CATHETER AND A THROMBUS ISSUE OCCURRED. DURING THE PROCEDURE, THROMBUS WAS DISCOVERED ON THE TIP OF THE THERMOCOOL® SMART TOUCH® SF BI-DIRECTIONAL NAVIGATION CATHETER. THE CATHETER WAS REPLACED, AND THE PROCEDURE WAS COMPLETED WITHOUT PATIENT CONSEQUENCE. MULTIPLE ATTEMPTS HAVE BEEN MADE TO OBTAIN CLARIFICATION TO THIS COMPLAINT. HOWEVER, NO FURTHER INFORMATION HAS BEEN MADE AVAILABLE. THE THROMBUS ON THE TIP OF THE CATHETER WAS ASSESSED AS A REPORTABLE ISSUE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1074697 THERMOCOOL® SMART TOUCH® SF BI-DIRECTIONAL NAVIGATION CATHETER CARDIAC ABLATION PERCUTANEOUS CATHETER LPB BIOSENSE WEBSTER INC 30255929M 10846835010183

Patients

Seq Age Sex Outcome Treatment
1 UNK_SMARTABLATE GENERATOR.