FDA Adverse Event Malfunction Summary report: N

APPLICATION INSTRUMENT FOR STERNAL ZIPFIX

MDR report key: 9280357 · Received November 5, 2019

Report

Report Number
2939274-2019-61980
Event Type
Malfunction
Date Received
November 5, 2019
Date of Event
October 18, 2019
Report Date
October 18, 2019
Manufacturer
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
Product Code
JDQ
UDI-DI
10887587010847
PMA / PMN Number
K110789
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 0

DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. DEVICE HISTORY LOT : PART: 03.501.080-US. LOT: 9812895. MANUFACTURING SITE: HÄGENDORF. RELEASE TO WAREHOUSE DATE: 05. FEB. 2016. A MANUFACTURING RECORD EVALUATION WAS PERFORMED FOR THE FINISHED DEVICE LOT NUMBER, AND NO NON-CONFORMANCES WERE IDENTIFIED. SERVICE EVALUATION: THE CUSTOMER REPORTED THE DEVICE WOULD NOT PROPERLY TIGHTEN. THE REPAIR TECHNICIAN REPORTED THE DEVICE WAS NOT IN SPECIFICATION PER THE INSPECTION SHEET. THE CAUSE OF THE ISSUE IS UNKNOWN. THE ITEM WILL BE REPAIRED PER THE INSPECTION SHEET AND UPON PASSING SYNTHES FINAL INSPECTION, WILL BE RETURNED TO THE CUSTOMER UPON COMPLETION OF THE SERVICE AND REPAIR PROCESS. THE EVALUATION WAS CONFIRMED. THE DEVICE WAS DEEMED SERVICEABLE AND WILL BE RETURNED TO THE CUSTOMER, NO DESIGN OR MANUFACTURING ISSUES WERE IDENTIFIED THEREFORE IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE

Additional Manufacturer Narrative · 0

DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. THE CUSTOMER REPORTED THE DEVICE WOULD NOT PROPERLY TIGHTEN. THE REPAIR TECHNICIAN REPORTED THE DEVICE WAS NOT IN SPECIFICATION PER THE INSPECTION SHEET. THE CAUSE OF THE ISSUE IS UNKNOWN. THE ITEM WILL BE REPAIRED PER THE INSPECTION SHEET AND UPON PASSING SYNTHES FINAL INSPECTION, WILL BE RETURNED TO THE CUSTOMER UPON COMPLETION OF THE SERVICE AND REPAIR PROCESS. ATTACHED SERVICE RECORD ROUTER COMPLETED THROUGH OPERATION 10. FINALIZED SERVICE RECORD WILL BE ARCHIVED IN TUNGSTEN DOCUMENT MANAGEMENT SYSTEM. THE EVALUATION WAS CONFIRMED. THE DEVICE WAS DEEMED SERVICEABLE AND WILL BE RETURNED TO THE CUSTOMER, NO DESIGN OR MANUFACTURING ISSUES WERE IDENTIFIED THEREFORE IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. DEVICE HISTORY LOT PART: 03.501.080-US. LOT: 9812895. MANUFACTURING SITE: HÄGENDORF. RELEASE TO WAREHOUSE DATE: 05. FEB. 2016. A MANUFACTURING RECORD EVALUATION WAS PERFORMED FOR THE FINISHED DEVICE LOT NUMBER, AND NO NON-CONFORMANCES WERE IDENTIFIED. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS.

Additional Manufacturer Narrative · 1

THE DEVICE WAS RECEIVED, THE INVESTIGATION IS IN PROGRESS, NO CONCLUSION COULD BE DRAWN AT THE TIME OF FILING THIS REPORT. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING AN UNKNOWN SURGERY ON (B)(6) 2019, AN APPLICATION INSTRUMENT FOR STERNAL ZIPFIX DID NOT PROPERLY TIGHTEN AN UNKNOWN ZIPFIX IMPLANT. THE SURGEON SIMPLY PASSED THE APPLICATOR OFF THE STERILE FIELD AND USED ANOTHER APPLICATOR. IT IS UNKNOWN IF THERE WAS A SURGICAL DELAY. THE SURGERY WAS COMPLETED WITH NO ADVERSE CONSEQUENCE TO THE PATIENT. CONCOMITANT DEVICES REPORTED: UNKNOWN ZIPFIX IMPLANT (PART# UNKNOWN, LOT# UNKNOWN, QUANTITY# 1). THIS REPORT IS FOR 1 OF 1 FOR (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1073174 APPLICATION INSTRUMENT FOR STERNAL ZIPFIX CERCLAGE FIXATION JDQ WRIGHTS LANE SYNTHES USA PRODUCTS LLC 03.501.080 10887587010847

Patients

Seq Age Sex Outcome Treatment
1 UNK - ZIPFIX IMPLANTS