LINER CANNSTR SUCT 3000CC DISP
Report
- Report Number
- 1423507-2007-00109
- Event Type
- Other
- Date Received
- October 15, 2007
- Date of Event
- September 27, 2007
- Report Date
- October 15, 2007
- Manufacturer
- CARDINAL HEALTH
- Product Code
- KDQ
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- LA, US
- Reporter Occupation
- OTHER
Narratives
THE ASSEMBLY RETURNED FOR INVESTIGATION EXHIBITED SEVERE DAMAGE INDICATIVE OF AN IMPLOSION EVENT WHERE THE LID WAS TOTALLY SEPARATED INTO MULTIPLE PIECES. THE LID REMNANT HELPED IDENTIFY THE LID CAVITY NUMBER AS "6". THE LID DATE CODE WAS 8-07 (AUGUST 2007), WHICH REPRESENTS THE DATE OF MANUFACTURE OF THE LID. IT WOULD BE EXPECTED THAT THE ASSEMBLY AND PACK OUT TIME FRAME WOULD ALSO BE WITHIN THIS GENERAL PERIOD. THE RETURNED UNIT WAS DIMENSIONALLY INSPECTED. THE LID COMPONENT MET THE SPECIFICATION REQUIREMENT FOR WALL THICKNESS. THE PRODUCT IS DESIGNED TO MEET IMPLOSION RESISTANCE REQUIREMENTS PER ISO 10079-3 "MEDICAL SUCTION EQUIPMENT- PART 3: SUCTION EQUIPMENT POWERED FROM VACUUM OR PRESSURE SOURCE. UNFORTUNATELY, NO LOT NUMBER WAS PROVIDED BY THE CUSTOMER; THEREFORE, AN INVESTIGATION OF THE MANUFACTURING DEVICE HISTORY RECORD COULD NOT BE PERFORMED. DISCUSSIONS WERE HELD WITH KEY QUALITY, MARKETING AND TECHNICAL PERSONNEL REGARDING THIS CONCERN, WHILE NO DEFINITIVE ROOT CAUSE COULD BE IDENTIFIED BY THE COMPLAINT INVESTIGATION, SUCH AN EVENT MAY OCCUR IF THE UNIT WAS DAMAGED PRIOR TO USE OR, IF THE PRODUCT WAS SUBJECT TO STATIC VACUUM PRIOR TO USE FOR AN EXTENDED PERIOD.
CANISTER LINER "EXPLODED" PER A CRNA WHO WAS THE ONLY PERSON IN THE OR AT THE TIME . WHEN THE EXPLOSION HAPPENED, THE LID SHATTERED, WENT PROJECTILE, AND LINER SPLIT DOWN THE SIDE. NO ONE WAS HANDLING THE CANISTER AT THE TIME.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LINER CANNSTR SUCT 3000CC DISP | SUCTION CANISTER LINER | KDQ | CARDINAL HEALTH | 65651-530 | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA YR |