FDA Adverse Event Other Summary report: N

LINER CANNSTR SUCT 3000CC DISP

MDR report key: 927850 · Received October 15, 2007

Report

Report Number
1423507-2007-00109
Event Type
Other
Date Received
October 15, 2007
Date of Event
September 27, 2007
Report Date
October 15, 2007
Manufacturer
CARDINAL HEALTH
Product Code
KDQ
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
LA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE ASSEMBLY RETURNED FOR INVESTIGATION EXHIBITED SEVERE DAMAGE INDICATIVE OF AN IMPLOSION EVENT WHERE THE LID WAS TOTALLY SEPARATED INTO MULTIPLE PIECES. THE LID REMNANT HELPED IDENTIFY THE LID CAVITY NUMBER AS "6". THE LID DATE CODE WAS 8-07 (AUGUST 2007), WHICH REPRESENTS THE DATE OF MANUFACTURE OF THE LID. IT WOULD BE EXPECTED THAT THE ASSEMBLY AND PACK OUT TIME FRAME WOULD ALSO BE WITHIN THIS GENERAL PERIOD. THE RETURNED UNIT WAS DIMENSIONALLY INSPECTED. THE LID COMPONENT MET THE SPECIFICATION REQUIREMENT FOR WALL THICKNESS. THE PRODUCT IS DESIGNED TO MEET IMPLOSION RESISTANCE REQUIREMENTS PER ISO 10079-3 "MEDICAL SUCTION EQUIPMENT- PART 3: SUCTION EQUIPMENT POWERED FROM VACUUM OR PRESSURE SOURCE. UNFORTUNATELY, NO LOT NUMBER WAS PROVIDED BY THE CUSTOMER; THEREFORE, AN INVESTIGATION OF THE MANUFACTURING DEVICE HISTORY RECORD COULD NOT BE PERFORMED. DISCUSSIONS WERE HELD WITH KEY QUALITY, MARKETING AND TECHNICAL PERSONNEL REGARDING THIS CONCERN, WHILE NO DEFINITIVE ROOT CAUSE COULD BE IDENTIFIED BY THE COMPLAINT INVESTIGATION, SUCH AN EVENT MAY OCCUR IF THE UNIT WAS DAMAGED PRIOR TO USE OR, IF THE PRODUCT WAS SUBJECT TO STATIC VACUUM PRIOR TO USE FOR AN EXTENDED PERIOD.

Description of Event or Problem · 1

CANISTER LINER "EXPLODED" PER A CRNA WHO WAS THE ONLY PERSON IN THE OR AT THE TIME . WHEN THE EXPLOSION HAPPENED, THE LID SHATTERED, WENT PROJECTILE, AND LINER SPLIT DOWN THE SIDE. NO ONE WAS HANDLING THE CANISTER AT THE TIME.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LINER CANNSTR SUCT 3000CC DISP SUCTION CANISTER LINER KDQ CARDINAL HEALTH 65651-530 UNK

Patients

Seq Age Sex Outcome Treatment
1 NA YR