FDA Adverse Event Malfunction Summary report: N

BD 5ML SYRINGE LUER-LOK TIP

MDR report key: 9273790 · Received November 4, 2019

Report

Report Number
1213809-2019-01103
Event Type
Malfunction
Date Received
November 4, 2019
Date of Event
October 7, 2019
Report Date
December 13, 2019
Manufacturer
BECTON DICKINSON MEDICAL SYSTEMS
Product Code
FMF
UDI-DI
30382903096498
PMA / PMN Number
K980987
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FI
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

H.6. INVESTIGATION SUMMARY: ONE PHOTO AND FOUR 5ML SYRINGES IN BLISTER PACKS FROM BATCH 9200385 (P/N 309649) WERE RECEIVED AND EVALUATED. THREE OF THE PACKS WERE FULLY SEALED AND ONE WAS OPENED. IT WAS OBSERVED ON ALL FOUR SYRINGES THERE WERE INK CONTACT SMEARS WITH EACH SYRINGE CONTAINING AT LEAST ONE INK SPOT LARGER THAN LEVEL 4 IN SIZE AND WERE REJECTABLE PER PRODUCT SPECIFICATION. POTENTIAL ROOT CAUSE FOR THE INK SMEARS DEFECT IS ASSOCIATED WITH THE MARKING PROCESS. IT IS POSSIBLE THE BARRELS MADE CONTACT SHORTLY AFTER BEING PRINTED. NO CORRECTIVE ACTIONS ARE NECESSARY BASED ON THE DEFECTIVE RATE IDENTIFIED. BATCH 9200385 IS CONSIDERED IN COMPLIANCE WITH OUR PRODUCT SPECIFICATION REQUIREMENTS. A DEVICE HISTORY RECORD REVIEW SHOWED NO REJECTED INSPECTIONS OR QUALITY ISSUES DURING THE PRODUCTION OF THE PROVIDED LOT NUMBER THAT COULD HAVE CONTRIBUTED TO THE REPORTED DEFECT. H3 OTHER TEXT : SEE SECTION H.10.

Description of Event or Problem · 0

IT WAS REPORTED THAT SCALE MARKING ISSUES WERE FOUND BEFORE USE WITH A BD 5ML SYRINGE LUER-LOK¿ TIP. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "IN THIS BATCH 4/13 PCE SAMPLE THERE IS PRINTING INK AROUND THE SYRINGES."

Additional Manufacturer Narrative · 1

(B)(6). A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT SCALE MARKING ISSUES WERE FOUND BEFORE USE WITH A BD 5ML SYRINGE LUER-LOK¿ TIP. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, "IN THIS BATCH 4/13 PCE SAMPLE THERE IS PRINTING INK AROUND THE SYRINGES."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1070804 BD 5ML SYRINGE LUER-LOK TIP PISTON SYRINGE FMF BECTON DICKINSON MEDICAL SYSTEMS 9200385 30382903096498

Patients

Seq Age Sex Outcome Treatment
1 Other