FDA Adverse Event Other Summary report: N

PROVOX BRUSH

MDR report key: 926115 · Received October 12, 2007

Report

Report Number
8032044-2007-00017
Event Type
Other
Date Received
October 12, 2007
Date of Event
August 26, 2006
Manufacturer
ATOS MEDICAL AB
Product Code
EWL
PMA / PMN Number
EXEMPTED
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

INVESTIGATION HAS REVEALED THAT THE PT HAS USED THE BRUSH FOR FIVE (5) YEARS. RECOMMENDATION IS 1 MONTH. THE PT RECEIVED ADDITIONAL INFORMATION. WE ALSO CONTINUOUSLY EDUCATE PTS, PROFESSIONALS AND SALES PERSONS OF THE CORRECT USE OF THE DEVICES. ALL MANUALS ARE ALSO DOWNLOADABLE FROM OUR WEBSITE. NO FURTHER ACTIONS. CASE HAS BEEN CLOSED. THIS EVENT WAS NOT INITIALLY REPORTED AS AN MDR. DUE TO INCIDENT REPORT IN 2007, (MFR#8032044-2007-00009) REVIEW ACCORDING TO OUR POSTMARKET SURVEILLANCE PROCEDURE OF FORMER SIMILAR EVENTS WAS DONE AND WE DECIDED TO REPORT THIS ISSUE.

Description of Event or Problem · 1

THE BRISTLE PART OF THE BRUSH BROKE OFF DURING CLEANING AND FELL DOWN IN THE PTS LUNG. THE PARTS WERE RETRIEVED DURING RADIOSCOPIC EXAMINATION. NO POSTOPERATIVE COMPLICATIONS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PROVOX BRUSH CLEANING BRUSH FOR VOICE PROSTHESIS EWL ATOS MEDICAL AB PROVOX BRUSH UNK

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention