FDA Adverse Event Malfunction Summary report: N

BD INSYTE¿ AUTOGUARD¿ BC SHIELDED IV CATHETER

MDR report key: 9257641 · Received October 30, 2019

Report

Report Number
1710034-2019-01159
Event Type
Malfunction
Date Received
October 30, 2019
Date of Event
October 8, 2019
Report Date
December 11, 2019
Manufacturer
BECTON DICKINSON INFUSION THERAPY SYSTEMS INC.
Product Code
FOZ
PMA / PMN Number
K110443
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
HK
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION: OUR QUALITY ENGINEER INSPECTED THE SAMPLE AND PICTURES PROVIDED. BD RECEIVED ONE UNUSED UNIT IN ITS ORIGINAL PACKAGING FROM LOT NUMBER 9071776. ONE PHOTO WAS ALSO RECEIVED. UPON VISUAL INSPECTION NO FOREIGN MATTER WAS OBSERVED ON THE UNIT. THROUGH MICROSCOPIC EVALUATION WE WERE ABLE TO CONFIRM FOREIGN MATTER ON THE CATHETER TUBING. THE FOREIGN MATTER APPEARS TO BE ON THE OUTSIDE OF THE CATHETER WALLS. THE UNIT WAS SENT FOR FURTHER FTIR TESTING. ONCE THE UNIT WAS RECEIVED FOR TESTING, FOREIGN MATTER WAS NO LONGER OBSERVED ON THE UNIT. UNFORTUNATELY, BD WAS UNABLE TO DETERMINE A ROOT CAUSE. A DEVICE HISTORY RECORD REVIEW SHOWED NO NON-CONFORMANCES ASSOCIATED WITH THIS ISSUE DURING THE PRODUCTION OF THIS BATCH.

Description of Event or Problem · 0

IT WAS REPORTED THAT FOREIGN MATTER WAS FOUND BEFORE USE WITH A BD INSYTE¿ AUTOGUARD¿ BC SHIELDED IV CATHETER. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, "THE USER COMPLAINED THAT BLACK FOREIGN BODY WAS FOUND AT THE TIP OF THE CATHETER."

Additional Manufacturer Narrative · 1

(B)(6). A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT FOREIGN MATTER WAS FOUND BEFORE USE WITH A BD INSYTE¿ AUTOGUARD¿ BC SHIELDED IV CATHETER. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, "THE USER COMPLAINED THAT BLACK FOREIGN BODY WAS FOUND AT THE TIP OF THE CATHETER."