FDA Adverse Event Other Summary report: N

MTS PIPETTE

MDR report key: 925327 · Received October 4, 2007

Report

Report Number
9681721-2007-00016
Event Type
Other
Date Received
October 4, 2007
Date of Event
August 10, 2007
Report Date
October 5, 2007
Manufacturer
BIOHIT OYJ
Product Code
JTC
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

SUMMARY: INDICATES THAT INCORRECT VOLUMES DISPENSED INVOLVING THE BIOHIT PIPETTOR CONSTITUTE A POTENTIAL HEALTH AND SAFETY HAZARD. INDICATES THAT AN INCORRECT DISPENSE, UNDETECTED BY THE USER (I.E. NO ERROR MESSAGE POSTED) IS A REPORTABLE INCIDENT. THIS CONDITION MAY LEAD TO ERRONEOUS TEST RESULTS. WE HAVE HERE A CUSTOMER RESPONDING CORRECTLY AND QUICKLY TO PREVENT FALSE READINGS AND INCORRECT TEST RESULTS. THE COMPLAINT WAS NEVER ACTUALLY CONFIRMED IN THIS INSTANCE, BUT THE PIPETTE WAS REPLACED AND NOT INTENDED FOR FURTHER USE AS IS. IT IS THE EXPERIENCE OF THE MANUFACTURER THAT THIS TYPE OF PROBLEM IS NOT ONLY CORRECTABLE ALLOWING THE PIPETTE TO AGAIN BE REUSED AFTER CORRECTIVE SERVICE, BUT IN MOST INSTANCES PREVENTABLE SHOULD THE CUSTOMER FOLLOW RECOMMENDED MANUFACTURER GUIDELINES FOR SCHEDULED PREVENTIVE SERVICE. BATTERY LIFE CAN BE EXTENDED BY FAITHFUL CHARGING HABITS TO PRECLUDE PREMATURE FAILURE OR REGULAR PREVENTIVE CHECKUPS AT OEM FACILITIES COULD DETECT A DAMAGED CHARGING SYSTEM OR FAILURE OF OTHER ELECTRONIC AND OR MECHANICAL COMPONENTS DUE TO AGEING AND THE NORMAL WEAR AND TEAR INCURRED BY USE.

Description of Event or Problem · 1

THE CUSTOMER INDICATED THAT THE BIOHIT PIPETTOR WAS SET TO PROGRAM 9, HOWEVER, AFTER ONE DISPENSE THE PIPETTOR FLASHED AND THEN SWITCHED TO PROGRAM 1 DURING TESTING. THE CUSTOMER DISCONTINUED THE USE OF THE PIPETTOR. THE CUSTOMER DID NOT PROVIDE ADDITIONAL INFORMATION REGARDING THIS INCIDENT. THE CUSTOMER INDICATED THAT THE BATTERY CONTACTS WERE CLEANED AND THE PIPETTOR WAS ALWAYS ON THE CHARGING STAND. CTS ADVISED THE CUSTOMER TO DISCONTINUE THE USE OF THE PIPETTOR AND TO CONTACT THEIR OCD ACCOUNT MANAGER REGARDING THE OPTION TO PURCHASE A NEW PIPETTOR. THE PIPETTOR WAS OUT OF WARRANTY. THE CUSTOMER WAS DISSATISFIED THAT OCD DOES NOT PROVIDE SERVICE FOR PIPETTORS THAT ARE OUT OF WARRANTY. NO ERRONEOUS RESULTS WERE REPORTED. A SEARCH WAS PERFORMED CONCERNING COMPLAINTS LOGGED BY THIS CUSTOMER. THIS CUSTOMER HAS NOT LOGGED ANY COMPLAINTS AGAINST THIS PIPETTOR SINCE THE INCIDENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MTS PIPETTE ELECTRONIC PIPETTOR JTC BIOHIT OYJ 710142MT NA

Patients

Seq Age Sex Outcome Treatment
1 YR