FDA Adverse Event
Injury
Summary report: N
GENTLEPOWER LUX CONTRA ANGLE 25LPA
MDR report key: 924922
·
Received October 10, 2007
Report
- Report Number
- 1419798-2007-00024
- Event Type
- Injury
- Date Received
- October 10, 2007
- Date of Event
- August 30, 2007
- Report Date
- September 7, 2007
- Manufacturer
- KAVO AMERICA
- Product Code
- EKK
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KS
- Reporter Occupation
- DENTIST
Narratives
Additional Manufacturer Narrative · 1
MAINTENANCE INFORMATION WAS REVIEWED WITH THE OFFICE AND MAINTENANCE SHEETS WERE SENT. THE WATER SPRAY WAS CLOGGED WITH A PRESENCE OF LOW RESIDUE. THE BACK CAP WAS STUCK IN DUE TO DENTS IN THE HEAD OF THE HANDPIECE.
Description of Event or Problem · 1
THE HEAD OF THE HANDPIECE BECAME HOT AND CREATED A SMALL BURN TO THE BUCCAL MUCOSA CHEEK. PATIENT WAS PRESCRIBED PERIDEX MOUTH RINSE AND HAS COMPLETELY HEALED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | GENTLEPOWER LUX CONTRA ANGLE 25LPA | OPERATIVE DENTAL UNIT ACCESSORY | EKK | KAVO AMERICA | 25LPA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | YR | Required Intervention |