FDA Adverse Event Injury Summary report: N

GENTLEPOWER LUX CONTRA ANGLE 25LPA

MDR report key: 924922 · Received October 10, 2007

Report

Report Number
1419798-2007-00024
Event Type
Injury
Date Received
October 10, 2007
Date of Event
August 30, 2007
Report Date
September 7, 2007
Manufacturer
KAVO AMERICA
Product Code
EKK
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
KS
Reporter Occupation
DENTIST

Narratives

Additional Manufacturer Narrative · 1

MAINTENANCE INFORMATION WAS REVIEWED WITH THE OFFICE AND MAINTENANCE SHEETS WERE SENT. THE WATER SPRAY WAS CLOGGED WITH A PRESENCE OF LOW RESIDUE. THE BACK CAP WAS STUCK IN DUE TO DENTS IN THE HEAD OF THE HANDPIECE.

Description of Event or Problem · 1

THE HEAD OF THE HANDPIECE BECAME HOT AND CREATED A SMALL BURN TO THE BUCCAL MUCOSA CHEEK. PATIENT WAS PRESCRIBED PERIDEX MOUTH RINSE AND HAS COMPLETELY HEALED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 GENTLEPOWER LUX CONTRA ANGLE 25LPA OPERATIVE DENTAL UNIT ACCESSORY EKK KAVO AMERICA 25LPA

Patients

Seq Age Sex Outcome Treatment
1 YR Required Intervention