FDA Adverse Event Injury Summary report: N

BLAZER II

MDR report key: 924921 · Received October 10, 2007

Report

Report Number
2953184-2007-00021
Event Type
Injury
Date Received
October 10, 2007
Report Date
September 17, 2007
Manufacturer
BOSTON SCIENTIFIC DP TECHNOLOGIES
Product Code
LPB
PMA / PMN Number
P020025
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

IT HAS BEEN ESTABLISHED THAT CONDITIONS THAT COULD LEAD TO AN UNINTENDED, RF INDUCED ARRYTHMIA WERE PRESENT DURING THE PROCEDURE. THIS MAY BE ATTRIBUTED TO AN IMPROPER EQUIPMENT COMBINATION FOR PACING WHILE ABLATING, AS IDENTIFIED IN A PREVIOUS BOSTON SCIENTIFIC STUDY. THE STUDY FOUND THAT CERTAIN COMBINATIONS OF STIMULATORS AND RECORDING SYSTEMS ARE SUSPECTABLE TO THE CREATION OF DC VOLTAGES ACROSS A PACING-ROUTED PAIR OF ELECTRODES DURING RF DELIVERY, WHICH CAN RESULT IN THE OCCURRENCE OF UNINTENDED CARDIAC STIMULATION. THESE CONDITIONS ARE DESCRIBED IN THE BSC WHITE PAPER "DC VOLTAGE GENERATION ACROSS DIAGNOSTIC ELECTRODES THROUGH INTERACTIONS BETWEEN EP RECORDING SYSTEMS, PACING STIMULATORS, AND RF GENERATORS." BOSTON SCIENTIFIC COMMUNICATED THIS INFORMATION IN A CUSTOMER LETTER TO ALL OF ITS EP CUSTOMERS IN AUGUST 2006. THERE WERE NO SIMILAR COMPLAINTS REPORTED FOR THIS DEVICE. ADD'L PMA/510(K) #P920047.

Description of Event or Problem · 1

IT WAS REPORTED TO BOSTON SCIENTIFIC THAT DURING AN ABLATION PROCEDURE WITH AN 8F BLAZER CATHETER, THE GROUNDING PADS WERE NOT PLUGGED IN WHILE USING A ST. JUDE NAV-X MAPPING SYSTEM. AFTER DELIVERING POWER FOR 25 SECONDS, THE PATIENT WENT INTO A-FIB WHICH TERMINATED ON ITS OWN. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH THIS CATHETER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 BLAZER II THERAPEUTIC ABLATION CATHETER LPB BOSTON SCIENTIFIC DP TECHNOLOGIES UNK UNK

Patients

Seq Age Sex Outcome Treatment
1 YR Other PRUCKA RECORDING SYSTEM| BSC XP GENERATOR AND APM| ST. JUDE NAVX MAPPING SYSTEM| BLOOM STIMULATOR