L SHAPED LSMI THREADLOCK DISTRACTOR, RIGHT
Report
- Report Number
- 1032347-2007-00027
- Event Type
- Other
- Date Received
- October 5, 2007
- Date of Event
- September 20, 2007
- Report Date
- September 20, 2007
- Manufacturer
- BIOMET MICROFIXATION, INC.
- Product Code
- JEY
- PMA / PMN Number
- K030425
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- PHYSICIAN
Narratives
THERE ARE WARNINGS IN THE PACKAGE INSERT THAT STATE THAT THIS TYPE OF EVENT CAN OCCUR. IFU INDICATES THE DEVICE CAN BREAK, BEND OR BE DAMAGED AS A RESULT OF STRESS, ACTIVITY, AND LOAD BEARING. CURRENT INFO IS INSUFFICIENT TO PERMIT A VALID CONCLUSION ABOUT THE CAUSE OF THIS EVENT. IF ADD'L INFO IS OBTAINED THAT ADDS VALUE TO THE RELEVANT CONTENT OF THIS REPORT AND/OR A CONCLUSION CAN BE DRAWN, A FOLLOW-UP REPORT WILL BE SENT. ADVISED USER FACILITY OF MEDWATCH FILING. 1032347-2007-00026 & 00027. REPORT FROM VISUAL REVIEW OF EXPLANTED PRODUCT. PHYSICIAN CONFIRMED (ALONG WITH POST OP PICTURES IN OR) THE EXTENSIONS WERE NOT HELD AGAINST THE PT'S HEAD AND ALLOWED TO DANGLE FREELY. ENGINEER BELIEVES IT IS LIKELY THE EXTENSIONS CAUGHT ON SOMETHING FORCING THEM LATERALLY AWAY FROM THE MANDIBLE (OR INFERIOR /SUPERIOR) TO CAUSE THE HOUSING TO BREAK. FURTHERMORE, ENGINEER NOTED THE EXTENSION OF THE PT'S RIGHT SIDE HAD BROKEN NEAR THE WELD CONNECTION BETWEEN THE THREADED SECTION AND FLEXIBLE EXTENSION.
MANDIBULAR DISTRACTION. DEVICE IMPLANTED IN 2007. THE NEXT MONTH, DOCTOR NOTICED ACTIVATION OF THE DRIVE SCREW WAS NOT ADVANCING BONE SEGMENTS. REVISION SURGERY FIVE DAYS LATER, DISCOVERED THE HOUSING OF THE LACTOSORB FOOT PLATE HAD BROKEN ALONG THE LENGTH OF THE INTERNAL THREADS OF THE PROXIMAL PLATE, WHERE THE HOUSING BLENDS INTO THE FLAT PORTION OF THE PLATE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | L SHAPED LSMI THREADLOCK DISTRACTOR, RIGHT | LACTOSORB DISTRACTION | JEY | BIOMET MICROFIXATION, INC. | 871450 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 1 YR | Required Intervention | FIXATE PLATES| (SEE 1032347-2007-0027). SCREWS WERE USED TO| LEFT AND RIGHT DISTRACTION PLATES WERE USED |