FDA Adverse Event Injury Summary report: N

M2A-TAPER LINER SZ 41/32

MDR report key: 9243728 · Received October 28, 2019

Report

Report Number
0001825034-2019-04826
Event Type
Injury
Date Received
October 28, 2019
Date of Event
August 15, 2017
Report Date
September 8, 2022
Manufacturer
ZIMMER BIOMET, INC.
Product Code
JDL
PMA / PMN Number
K042841
Removal / Correction Number
N/A
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AL, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THIS FOLLOW-UP REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION. THE FOLLOWING SECTIONS WERE UPDATED/CORRECTED UPDATED: A1; A2; B5; B7; G3; H2; H6 IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS.

Additional Manufacturer Narrative · 0

THIS FOLLOW-UP REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION. MEDICAL RECORDS/RADIOGRAPHS WERE PROVIDED AND REVIEWED BY A HEALTH CARE PROFESSIONAL. A REVIEW OF THE AVAILABLE RECORDS IDENTIFIED FINDINGS OF THE REPORTED ISSUES, POST LEFT HIP ARTHROPLASTY WITH ELEVATED COBALT LEVELS FROM MOM WEAR, TISSUE NECROSIS DEBRIDED HIP. NO OTHER COMPLICATIONS NOTED. THE ADDITIONAL INFORMATION DOES NOT CHANGE THE OUTCOME OF THE PREVIOUS INVESTIGATION. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS.

Description of Event or Problem · 0

IT WAS REPORTED THAT PATIENT UNDERWENT A LEFT HIP REVISION APPROXIMATELY 16 YEARS POST IMPLANTATION DUE TO ELEVATED METAL ION LEVELS. DURING THE PROCEDURE, THERE WAS NOTED TISSUE NECROSIS. THE HEAD LINER AND SHELL WERE REMOVED AND REPLACED. ATTEMPTS HAVE BEEN MADE AND ADDITIONAL INFORMATION ON THE REPORTED EVENT IS UNAVAILABLE AT THIS TIME.

Description of Event or Problem · 0

NO ADDITIONAL EVENT INFORMATION TO REPORT AT THIS TIME.

Additional Manufacturer Narrative · 1

(B)(4). CONCOMITANT MEDICAL PRODUCTS: 15-104096 ¿ M2A TAPER SHELL ¿ 099440. 11-163673 ¿ M2A HEAD ¿ 452530. 11-112116 ¿IMPACT DISTAL ¿ 778100. 112085 ¿ METAPHYSEAL - 423090. REPORTED EVENT WAS UNABLE TO BE CONFIRMED DUE TO LIMITED INFORMATION RECEIVED FROM THE CUSTOMER. DHR WAS REVIEWED AND NO DISCREPANCIES WERE FOUND. ROOT CAUSE WAS UNABLE TO BE DETERMINED AS THE NECESSARY INFORMATION TO ADEQUATELY INVESTIGATE THE REPORTED EVENT WAS NOT PROVIDED. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS. MULTIPLE MDR REPORTS WERE FILED FOR THIS EVENT, PLEASE SEE ASSOCIATED REPORTS: 0001825034 - 2019 - 03037.

Description of Event or Problem · 1

IT WAS REPORTED THAT PATIENT UNDERWENT A LEFT HIP REVISION APPROXIMATELY 16 YEARS POST IMPLANTATION DUE TO PAIN, DISCOMFORT, LACK OF MOBILITY AND ELEVATED METAL ION LEVELS. ATTEMPTS HAVE BEEN MADE AND ADDITIONAL INFORMATION ON THE REPORTED EVENT IS UNAVAILABLE AT THIS TIME.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1041084 M2A-TAPER LINER SZ 41/32 PROSTHESIS, HIP JDL ZIMMER BIOMET, INC. N/A 060920

Patients

Seq Age Sex Outcome Treatment
1 Male Hospitalization| R