M2A-TAPER LINER SZ 41/32
Report
- Report Number
- 0001825034-2019-04826
- Event Type
- Injury
- Date Received
- October 28, 2019
- Date of Event
- August 15, 2017
- Report Date
- September 8, 2022
- Manufacturer
- ZIMMER BIOMET, INC.
- Product Code
- JDL
- PMA / PMN Number
- K042841
- Removal / Correction Number
- N/A
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AL, US
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
THIS FOLLOW-UP REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION. THE FOLLOWING SECTIONS WERE UPDATED/CORRECTED UPDATED: A1; A2; B5; B7; G3; H2; H6 IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS.
THIS FOLLOW-UP REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION. MEDICAL RECORDS/RADIOGRAPHS WERE PROVIDED AND REVIEWED BY A HEALTH CARE PROFESSIONAL. A REVIEW OF THE AVAILABLE RECORDS IDENTIFIED FINDINGS OF THE REPORTED ISSUES, POST LEFT HIP ARTHROPLASTY WITH ELEVATED COBALT LEVELS FROM MOM WEAR, TISSUE NECROSIS DEBRIDED HIP. NO OTHER COMPLICATIONS NOTED. THE ADDITIONAL INFORMATION DOES NOT CHANGE THE OUTCOME OF THE PREVIOUS INVESTIGATION. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS.
IT WAS REPORTED THAT PATIENT UNDERWENT A LEFT HIP REVISION APPROXIMATELY 16 YEARS POST IMPLANTATION DUE TO ELEVATED METAL ION LEVELS. DURING THE PROCEDURE, THERE WAS NOTED TISSUE NECROSIS. THE HEAD LINER AND SHELL WERE REMOVED AND REPLACED. ATTEMPTS HAVE BEEN MADE AND ADDITIONAL INFORMATION ON THE REPORTED EVENT IS UNAVAILABLE AT THIS TIME.
NO ADDITIONAL EVENT INFORMATION TO REPORT AT THIS TIME.
(B)(4). CONCOMITANT MEDICAL PRODUCTS: 15-104096 ¿ M2A TAPER SHELL ¿ 099440. 11-163673 ¿ M2A HEAD ¿ 452530. 11-112116 ¿IMPACT DISTAL ¿ 778100. 112085 ¿ METAPHYSEAL - 423090. REPORTED EVENT WAS UNABLE TO BE CONFIRMED DUE TO LIMITED INFORMATION RECEIVED FROM THE CUSTOMER. DHR WAS REVIEWED AND NO DISCREPANCIES WERE FOUND. ROOT CAUSE WAS UNABLE TO BE DETERMINED AS THE NECESSARY INFORMATION TO ADEQUATELY INVESTIGATE THE REPORTED EVENT WAS NOT PROVIDED. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS. MULTIPLE MDR REPORTS WERE FILED FOR THIS EVENT, PLEASE SEE ASSOCIATED REPORTS: 0001825034 - 2019 - 03037.
IT WAS REPORTED THAT PATIENT UNDERWENT A LEFT HIP REVISION APPROXIMATELY 16 YEARS POST IMPLANTATION DUE TO PAIN, DISCOMFORT, LACK OF MOBILITY AND ELEVATED METAL ION LEVELS. ATTEMPTS HAVE BEEN MADE AND ADDITIONAL INFORMATION ON THE REPORTED EVENT IS UNAVAILABLE AT THIS TIME.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1041084 | M2A-TAPER LINER SZ 41/32 | PROSTHESIS, HIP | JDL | ZIMMER BIOMET, INC. | N/A | 060920 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Male | Hospitalization| R |