FDA Adverse Event Other Summary report: N

PROVOX BRUSH

MDR report key: 924193 · Received October 2, 2007

Report

Report Number
8032044-2007-00015
Event Type
Other
Date Received
October 2, 2007
Date of Event
September 30, 2005
Report Date
October 30, 2007
Manufacturer
ATOS MEDICAL AB
Product Code
EWL
PMA / PMN Number
EXEMPTED
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

EXAMINATION HAVE NOT GIVEN ANY CONCLUSIVE EVIDENCE WHY THIS HAS HAPPENED. THE PATIENT HAVE MODIFIED THE BRUSHES BY SHORTENING THEM WITH A PAIR OF PLIERS. THIS MIGHT HAVE CAUSED THE WHITE END TO FALL OFF. PATIENT WAS INSTRUCTED NOT TO MODIFY THE BRUSHES AND THE PROBLEM SEEMS TO HAVE ENDED. CASE HAS BEEN CLOSED. THIS EVENT WAS NOT INITIALLY REPORTED AS AN MDR. DUE TO INCIDENT REPORT 06/29/2007, (MFR # 8032044-2007-00009) REVIEW ACCORDING TO OUR POSTMARKET SURVEILLANCE PROCEDURE OF FORMER SIMILAR EVENTS WAS DONE AND WE DECIDED TO REPORT THIS ISSUE.

Description of Event or Problem · 1

THE PATIENT USES APPROX. 1 BRUSH PER MONTH. AT SEVERAL OCCASIONS SHE HAS NOTICED THAT THE WHITE TIP HAS DISAPPEARED. PATIENT HAS NOT EXPERIENCED ANY DISCOMFORT AND DOES NOT KNOW WHEN THE TIP HAS DISAPPEARED. NO EFFECT ON THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PROVOX BRUSH CLEANING BRUSH FOR VOICE PROSTHESIS EWL ATOS MEDICAL AB PROVOX BRUSH 0412008

Patients

Seq Age Sex Outcome Treatment
1 YR