CATALYS SYSTEM
Report
- Report Number
- 3005675890-2019-00091
- Event Type
- Malfunction
- Date Received
- October 25, 2019
- Report Date
- October 28, 2020
- Manufacturer
- JOHNSON & JOHNSON SURGICAL VISION, INC.
- Product Code
- OOE
- UDI-DI
- 05050474609792
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
CORRECTION: IT WAS INITIALLY REPORTED THAT 2 OUT OF 3 INVESTIGATIONS WERE COMPLETED DURING THE PERIOD HOWEVER, ONLY ONE INVESTIGATION WAS COMPLETED. ADDITIONAL INFORMATION TWO INVESTIGATIONS WERE COMPLETED DURING THIS PERIOD. A REVIEW OF RECORDS RELATED TO THE DEVICE INCLUDING LABELING, COMPLAINT TRENDING, AND RISK DOCUMENTATION WAS PERFORMED. THERE WERE NO ISSUES FOUND WITH THE SYSTEM. ALL PERTINENT INFORMATION AVAILABLE TO JOHNSON & JOHNSON SURGICAL VISION, INC HAS BEEN SUBMITTED.
THIS CORRECTION MDR IS BEING SUBMITTED TO ADDRESS INACCURATE INFORMATION IN SECTION F AND G. 4 DATE ON THE XML VERSION OF THE PREVIOUS SUBMISSION CAUSED BY A SOFTWARE GLITCH IN THE TRANSMISSION PROCESS. THE INTERNAL NON-CONFORMANCE NUMBERS FOR THIS SOFTWARE ISSUE ARE NR-0148919 AND CAPA-010215.
ADDITIONAL INFORMATION SERIAL NUMBERS OF DEVICES: (B)(4). PMA/510K: THIS REPORT IS BEING FILED ON AN INTERNATIONAL PRODUCT; LASER CORRECTION SYSTEM, MODEL CATALYS-I, THAT HAS A SIMILAR PRODUCT, CATALYS-U WHICH IS DISTRIBUTED IN THE UNITED STATES UNDER 510(K) # K121091. THE CATALYS SYSTEM MAY HAVE A REMOTE POTENTIAL EVENT WHERE LOSS OF SUCTION DURING LASER FIRING COULD CREATE A HAZARDOUS SITUATION THAT MAY RESULT IN SCORING OF THE POSTERIOR CORNEAL SURFACE. THEREFORE JOHNSON & JOHNSON SURGICAL VISION, INC HAS ISSUED AN ADVISORY NOTICE TO REINFORCE INSTRUCTIONS PROVIDED IN PRODUCT TRAINING AND IN THE OPERATOR¿S MANUAL REGARDING SUCTION LOSS DURING A PROCEDURE. INVESTIGATION WERE COMPLETED FOR 2 OUT 3 OF THE REPORTED EVENTS. A REVIEW OF RECORDS RELATED TO THE DEVICE INCLUDING LABELING, COMPLAINT TRENDING, AND RISK DOCUMENTATION WAS PERFORMED. THERE WERE NO ISSUES FOUND WITH THE SYSTEM. A REVIEW OF RECORDS RELATED TO THE DEVICE INCLUDING LABELING, COMPLAINT TRENDING, AND RISK DOCUMENTATION WILL BE PERFORMED. UPON COMPLETION OF THIS REVIEW, IF THERE IS ANY FURTHER RELEVANT INFORMATION OBTAINED, A SUPPLEMENTAL MEDWATCH WILL BE FILED. ALL PERTINENT INFORMATION AVAILABLE TO JOHNSON & JOHNSON SURGICAL VISION, INC. THAT WAS RECEIVED DURING THE TIME PERIOD THAT COVERS THIS VOLUNTARY MALFUNCTION SUMMARY REPORT HAS BEEN SUBMITTED.
THIS REPORT SUMMARIZES <NOE> 3 </NOE> MALFUNCTION EVENTS. THE EVENTS WERE RELATED TO SUCTION LOSS WHILE THE CATALYS LASER WAS FIRING. THERE WERE NO MEDICAL EVENTS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1031694 | CATALYS SYSTEM | CATALYS | OOE | JOHNSON & JOHNSON SURGICAL VISION, INC. | CATALYS-I | 05050474609792 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |