FDA Adverse Event Malfunction Summary report: N

BD VACUTAINER® C&S PRESERVATIVE URINE TUBE

MDR report key: 9240212 · Received October 25, 2019

Report

Report Number
1917413-2019-02254
Event Type
Malfunction
Date Received
October 25, 2019
Date of Event
October 10, 2019
Report Date
November 7, 2019
Manufacturer
BECTON, DICKINSON & CO. (BROKEN BOW)
Product Code
JSM
UDI-DI
50382903649513
PMA / PMN Number
K024240
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION SUMMARY: BD RECEIVED SAMPLES AND PHOTOS FROM THE CUSTOMER FACILITY FOR INVESTIGATION. THE PHOTOS AND SAMPLES WERE EVALUATED AND IT WAS OBSERVED THAT THE ADDITIVE WAS IN LIQUID FORM. A REVIEW OF THE DEVICE HISTORY RECORD WAS COMPLETED FOR THE INCIDENT LOT AND, BASED ON THIS REVIEW, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET; THERE WERE NO RELATED QUALITY NON-CONFORMANCES DURING MANUFACTURING OF THE PRODUCT.

Description of Event or Problem · 0

IT WAS REPORTED THAT AN UNSPECIFIED NUMBER OF BD VACUTAINER® C&S PRESERVATIVE URINE TUBES EXPERIENCED DISCOLORED/ABNORMAL ADDITIVE FORM WHICH WAS NOTED PRIOR TO USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: MATERIAL NO. 364951, BATCH NO. 9220404. PER EMAIL: WE JUST RECEIVED 3 CASES OF 364951 URINE C&S PRESERVATIVE. THE TUBES CAME WITH THE WHITE PRESERVATIVE IN LIQUID FORM INSTEAD OF POWDER.

Additional Manufacturer Narrative · 1

A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT AN UNSPECIFIED NUMBER OF BD VACUTAINER® C&S PRESERVATIVE URINE TUBES EXPERIENCED DISCOLORED/ABNORMAL ADDITIVE FORM WHICH WAS NOTED PRIOR TO USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: MATERIAL NO. 364951, BATCH NO. 9220404. PER EMAIL: WE JUST RECEIVED 3 CASES OF 364951 URINE C&S PRESERVATIVE. THE TUBES CAME WITH THE WHITE PRESERVATIVE IN LIQUID FORM INSTEAD OF POWDER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1037095 BD VACUTAINER® C&S PRESERVATIVE URINE TUBE TRANSPORT CULTURE MEDIUM JSM BECTON, DICKINSON & CO. (BROKEN BOW) 9220404 50382903649513

Patients

Seq Age Sex Outcome Treatment
1 Other