INRATIO
Report
- Report Number
- 2954730-2007-00479
- Event Type
- Malfunction
- Date Received
- October 4, 2007
- Date of Event
- September 21, 2007
- Report Date
- October 4, 2007
- Manufacturer
- HEMOSENSE, INC.
- Product Code
- JPA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- UNKNOWN
Narratives
INRATIO PRECISION DATA PROVIDED BY END-USER LOT: 070274/070308. FIRST TEST INR = 1.5, SECOND TEST INR = 1.3; MEAN = 1.9; SD = 0.57; %CV = 29.8%. THE %CV IS GREATER THAN 20%. PER INTERNAL PROCEDURE, THE PRECISION FAILS THE CRITERIA FOR PRECISION. THE TEST IS CONSIDERED NOT PRECISE AND FURTHER TESTING IS REQUIRED AT THIS TIME. DISCREPANT RESULTS (ACCURACY) COMPARISON OF INRATIO TEST WITH LAB RESULTS PROVIDED BY END-USER AT TIME COMPLAINT WAS FILED: PER INTERNAL PROCEDURE, THE MEAN OF THE INRATIO METER AND COMPARATIVE SYSTEM INR WERE CALCULATED. BOTH INRATIO AND LAB VALUES ARE WITHIN THE CONFIDENCE LIMITS FOR INR TESTING FOR ALL DATA SETS. THE RESULTS ARE NOT CONSIDERED DISCREPANT WITHIN THE CONTEXT OF THE DOCUMENTED VARIABILITY FOR INR TESTING. THEREFORE, FURTHER TESTING IS NOT REQUIRED AT THIS TIME.
CALLER ALLEGES IMPRECISION WITH INRATIO. RESULTS AS FOLLOWS: FIRST TEST INR = 1.5, SECOND TEST INR = 1.3. CALLER ALLEGES INACCURACY WITH INRATIO.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | INRATIO | PROTHROMBIN TIME TEST | JPA | HEMOSENSE, INC. | 070274 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | YR |