SLALOM PTA DILATATION CATHETER
Doing adverse event research by hand?
The Research Assistant searches MAUDE for you, groups reports by failure mode, and cites every record.
Free to try · every answer cites its records
Report
- Report Number
- 9610978-2007-00195
- Event Type
- Malfunction
- Date Received
- March 29, 2007
- Date of Event
- March 21, 2007
- Report Date
- March 21, 2007
- Manufacturer
- CORDIS EUROPA, N.V.
- Product Code
- LIT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
THE DEVICE WAS NOT RETURNED TO CORDIS FOR ANALYSIS. A DHR REVIEW WAS PERFORMED AND SHOWED THAT THIS LOT OF PRODUCTS MET ALL REQUIREMENTS PER THE APPLICABLE MQP. THIS REVIEW WAS EXTENDED TO SUBASSEMBLY PART NUMBER E7144621 WITH LOT NUMBER 0411066472. THIS REVIEW REVEALED THAT THE SUBASSEMBLIES MET ALL REQUIREMENTS PER THE APPLICABLE MQP AS WELL, INCLUDING BURST TEST VALUES. WITHOUT RETURN OF THE PRODUCT, IT CANNOT BE CONFIRMED IF THE OUTER SURFACE OF THE BALLOON MATERIAL REVEALED EVIDENCE OF SURFACE ABRASIONS THAT MIGHT HAVE CAUSED THE BALLOON BURST. WITH THE INFORMATION AVAILABLE, FACTORS THAT CONTRIBUTED TO THE REPORTED BALLOON BURST CANNOT BE DETERMINED.
IT WAS REPORTED THAT THE BALLOON RUPTURED. THE INTENDED PROCEDURE WAS ANGIOPLASTY OF A SHUNT ANASTOMOSIS. PER REPORT, THE VESSEL WAS NEITHER CALCIFIED NOR TORTUOUS, THE RATE OF STENOSIS IS UNK. THE SLALOM THRILL BALLOON CATHETER WAS USED TO DILATE THE LESION. DURING THE FOURTH INFLATION, THE BALLOON RUPTURED AT NOMINAL PRESSURE. AN INDEFLATOR WAS USED. NO ADDITIONAL PATIENT, LESION/VESSEL OR PROCEDURAL DETAILS WERE PROVIDED. IT IS UNK IF ANY DIFFICULTY WAS EXPERIENCED WHILE ADVANCING THE BALLOON CATHETER TO THE LESION OR WHILE CROSSING THE LESION. IT IS UNKNOWN IF THE BALLOON WAS INFLATED ABOVE RATED BURST PRESSURE ON ANY OR ALL OF THE PRIOR THREE INFLATIONS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SLALOM PTA DILATATION CATHETER | PTA CATHETERS | LIT | CORDIS EUROPA, N.V. | NA | R1206294 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 54 YR |