FDA Adverse Event Malfunction Summary report: N

SYRINGE HEPARIN 27GAX1/2 1ML

MDR report key: 9233622 · Received October 24, 2019

Report

Report Number
1920898-2019-01188
Event Type
Malfunction
Date Received
October 24, 2019
Date of Event
October 10, 2019
Report Date
October 28, 2019
Manufacturer
BD MEDICAL - DIABETES CARE
Product Code
FMF
PMA / PMN Number
N/A
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION: CUSTOMER RETURNED (3) LOOSE 1ML BD SYRINGE LABELED FOR CALCIPARINE FROM LOT # 7177608. CUSTOMER STATES THAT THE PRINTING IS THICKER THAN EXPECTED AND THERE IS A PRESENCE OF A LIQUID. ALL RETURNED SYRINGES WERE EXAMINED AND ALL SAMPLES EXHIBITED SMEARED SCALE MARKINGS PRINTED ON THE BARREL. HOWEVER, NO SAMPLE EXHIBITED ANY FOREIGN MATTER IN OR ON THE SAMPLES. A REVIEW OF THE DEVICE HISTORY RECORD WAS COMPLETED FOR BATCH # 7177608. ALL INSPECTIONS AND CHALLENGES WERE PERFORMED PER THE APPLICABLE OPERATIONS QC SPECIFICATIONS. THERE WERE ZERO (0) NOTIFICATIONS NOTED THAT PERTAINED TO THE COMPLAINT. UNABLE TO DETERMINE THE ROOT CAUSE OF THE SMEARED PRINT. CAPA#162566 WAS OPENED ON 25OCT2017 AFTER THE DATE OF MANUFACTURE TO ADDRESS PRINTING DEFECTS."

Description of Event or Problem · 0

IT WAS REPORTED THAT BEFORE USE OF THE SYRINGE HEPARIN 27GAX1/2 1ML THERE WERE SCALE MARKING ISSUES AND ISSUES WITH FOREIGN MATTER. THERE WERE 10 SYRINGES WITH THESE ISSUES. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM FRENCH TO ENGLISH: PRINTING THICKER THAN EXPECTED ON PLASTIC. PRESENCE OF LIQUID. 10 SYRINGES DETECTED IN 1 BOX. NO DATA ON THE WHOLE LOT.

Additional Manufacturer Narrative · 1

DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION AND/OR DEVICE HISTORY REVIEW, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT BEFORE USE OF THE SYRINGE HEPARIN 27GAX1/2 1ML THERE WERE SCALE MARKING ISSUES AND ISSUES WITH FOREIGN MATTER. THERE WERE 10 SYRINGES WITH THESE ISSUES. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM (B)(6) TO ENGLISH: PRINTING THICKER THAN EXPECTED ON PLASTIC. PRESENCE OF LIQUID. 10 SYRINGES DETECTED IN 1 BOX. NO DATA ON THE WHOLE LOT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1030890 SYRINGE HEPARIN 27GAX1/2 1ML SYRINGE FMF BD MEDICAL - DIABETES CARE 7177608

Patients

Seq Age Sex Outcome Treatment
1 Other