UNICEL DXC 800 PRO SYNCHRON CLINICAL SYSTEMS
Report
- Report Number
- 2050012-2007-00026
- Event Type
- Other
- Date Received
- October 4, 2007
- Date of Event
- August 29, 2007
- Report Date
- October 4, 2007
- Manufacturer
- BECKMAN COULTER, INC.
- Product Code
- JJE
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- OTHER
Narratives
NO QC PROBLEM WAS REPORTED. THE INSTRUMENT DID NOT GENERATE ERROR MESSAGES. A FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO THE CUSTOMER'S LAB: THE FSE REVIEWED THE EVENT AND DISCOVERED THAT THE INCORRECT CRP RESULTS WERE ALL RUN IN CUVETTE #6. PER FSE, NO OTHER ANALYTES WERE AFFECTED. THE FSE CONDUCTED A TEST TO USE ALL THE CUVETTES AND ONLY CUVETTE #6 FAILED TWICE. BASED ON OBTAINED DATA, THE FSE REALIZED A SIGNAL-PROCESSING ISSUE RATHER THAN CUVETTE ISSUE. A SMART MODULE 81 (PHOTOMETER ANALOG) WAS REPLACED WHICH RESOLVED THE PROBLEM. THE SMART MODULE HAS BEEN RETURNED TO BCI FOR INVESTIGATION. A HARDWARE ISSUE, ADDRESSED BY THE FSE, MAY HAVE CONTRIBUTED TO THIS EVENT. A MALFUNCTION WILL BE ASSUMED FOR THE PURPOSE OF THIS REPORT.
A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) REGARDING FALSELY LOW C-REACTIVE PROTEIN (CRP) RESULTS GENERATED BY THE UNICEL DXC 800 SYNCHRON CLINICAL SYSTEMS FOR FOUR PATIENTS. PATIENT A: SPECIMENS WERE TESTED BETWEEN TWO DAYS IN 2007. CRP RESULTS WERE IN THE RANGE OF "<5MG/DL" TO 174MG/DL. PATIENT B: SPECIMENS WERE TESTED IN 2007. CRP RESULTS WERE IN THE RANGE OF "<5MG/DL" TO 202MG/DL. HISTORICAL RESULTS FOR PATIENT B WERE: 218MG/DL AND 194MG/DL. PATIENT C: SPECIMENS WERE TESTED IN 2007. CRP RESULTS WERE IN THE RANGE OF "<5MG/DL" TO 317MG/DL. HISTORICAL RESULTS FOR PATIENT C WERE IN THE RANGE OF 88MG/DL-240MG/DL. PATIENT D: SPECIMENS WERE TESTED IN 2007. CRP RESULTS WERE IN THE RANGE OF "5MG/DL"TO 223MG/DL. PATIENT TREATMENT WAS NOT AFFECTED UPON RESULTS OBTAINED IN THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | UNICEL DXC 800 PRO SYNCHRON CLINICAL SYSTEMS | CLINICAL CHEMISTRY ANALYZER | JJE | BECKMAN COULTER, INC. | DXC 800 PRO | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA YR | Other |