FDA Adverse Event Malfunction Summary report: N

HANDPIECE MICS

MDR report key: 9227696 · Received October 23, 2019

Report

Report Number
3005985723-2019-00752
Event Type
Malfunction
Date Received
October 23, 2019
Date of Event
October 16, 2019
Report Date
January 28, 2020
Manufacturer
MAKO SURGICAL CORP.
Product Code
OLO
UDI-DI
00848486030193
PMA / PMN Number
K142530
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

TRIGGER TO UNLOCK HANDLE TO ROTATE, THE PIN FELL OUT. HANDLE LOCKED IN PLACE. CANNOT SEE LOT NUMBER BECAUSE OF THAT. CASE TYPE: TKA. SURGICAL DELAY: =15 MINUTES. PRODUCT EVALUATION AND RESULTS: MICS-209063, SN# (B)(6), RMA#282620. INSPECTED PER D06917 AND DETERMINED FAILURE OF THE FOLLOWING TEST STEP. SEC# 7.1.2. VISUAL PIVOT HANDLE. THE PIVOT HANDLE LOCK WAS BROKEN OFF COULD NOT ROTATE BUT I WAS ABLE TO VERIFY THE SN# (B)(6). DISPOSITION: RTV INSPECTED BY: (B)(6). PRODUCT HISTORY REVIEW: DEVICE HISTORY RECORDS INDICATE 25 DEVICES WERE MANUFACTURED UNDER LOT 42040317 AND 24 DEVICES WERE ACCEPTED INTO FINAL STOCK ON 28 APRIL, 2017. A REVIEW OF QT17-04-0080 REVEALED THAT THE ISSUE IS NOT RELATED TO THE FAILURE ALLEGED IN THIS COMPLIANT. COMPLAINT HISTORY REVIEW: A REVIEW OF COMPLAINTS IN CATSWEB AND TRACKWISE RELATED TO P/N 209063, L/N: 42040317 SHOWS 0 ADDITIONAL COMPLAINTS RELATED TO THE FAILURE IN THIS INVESTIGATION. CONCLUSIONS: THE ALLEGED FAILURE MODE WAS CONFIRMED. NO ADDITIONAL INVESTIGATION OR SPECIFIC ACTIONS ARE REQUIRED.

Description of Event or Problem · 0

TRIGGER TO UNLOCK HANDLE TO ROTATE, THE PIN FELL OUT. HANDLE LOCKED IN PLACE. CANNOT SEE LOT NUMBER BECAUSE OF THAT. CASE TYPE: TKA. SURGICAL DELAY: =15 MINUTES.

Additional Manufacturer Narrative · 1

AS PART OF THE NORMAL COMPLAINT FOLLOW-UP, AN EVALUATION OF THE EVENT HAS BEEN INITIATED BY MAKO SURGICAL. A SUPPLEMENTAL REPORT WILL BE SUBMITTED WHEN ADDITIONAL INFORMATION BECOMES AVAILABLE.

Description of Event or Problem · 1

TRIGGER TO UNLOCK HANDLE TO ROTATE, THE PIN FELL OUT. HANDLE LOCKED IN PLACE. CANNOT SEE LOT NUMBER BECAUSE OF THAT. CASE TYPE: TKA. SURGICAL DELAY: =15 MINUTES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1020864 HANDPIECE MICS STEREOTAXIC DEVICE, ROBOTICS OLO MAKO SURGICAL CORP. 209063 42040317 00848486030193

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization