FDA Adverse Event Injury Summary report: N

BOLT F/FEM NECK SYST F/CONSTRUCT LENGTH

MDR report key: 9227262 · Received October 23, 2019

Report

Report Number
8030965-2019-69582
Event Type
Injury
Date Received
October 23, 2019
Report Date
October 7, 2019
Manufacturer
OBERDORF SYNTHES PRODUKTIONS GMBH
Product Code
KTT
PMA / PMN Number
K172872
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 0

DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF THE INFORMATION IS UNKNOWN, NOT AVAILABLE OR DOES NOT APPLY, THE SECTION/FIELD OF THE FORM IS LEFT BLANK. H10: ADDITIONAL NARRATIVE. B6: TEST DATA UPDATED. H3, H6: INVESTIGATION SUMMARY: THE COMPLAINT CONDITION CAN BE CONFIRMED ACCORDING TO THE RECEIVED PICTURES. PRODUCT WAS NOT RETURNED FOR INVESTIGATION. BASED ON THE INFORMATION AVAILABLE, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND PREVENTATIVE ACTION IS PROPOSED. THIS COMPLAINT WILL BE ACCOUNTED FOR AND MONITORED VIA POST MARKET SURVEILLANCE ACTIVITIES. IF ADDITIONAL INFORMATION IS MADE AVAILABLE, THE INVESTIGATION WILL BE UPDATED AS APPLICABLE. DEVICE HISTORY LOT . PART NUMBER: 04.168.285S, SYNTHES LOT NUMBER: L873526, MANUFACTURING SITE: GRENCHEN, RELEASE TO WAREHOUSE DATE: APRIL 26, 2018, EXPIRY DATE: APRIL 01, 2028. A MANUFACTURING RECORD EVALUATION WAS PERFORMED FOR THE FINISHED DEVICE LOT NUMBER, AND NO NON-CONFORMANCES WERE IDENTIFIED. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 0

DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 0

DATE OF REVISION: (B)(6) 2019.

Additional Manufacturer Narrative · 1

COMPLAINANT PART IS NOT EXPECTED TO BE RETURNED FOR MANUFACTURER REVIEW/INVESTIGATION. (B)(4). THE INVESTIGATION COULD NOT BE COMPLETED; NO CONCLUSION COULD BE DRAWN, AS NO PRODUCT WAS RECEIVED. BASED ON THE INFORMATION AVAILABLE, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. THIS COMPLAINT WILL BE ACCOUNTED FOR AND MONITORED VIA POST MARKET SURVEILLANCE ACTIVITIES. IF ADDITIONAL INFORMATION IS MADE AVAILABLE, THE INVESTIGATION WILL BE UPDATED AS APPLICABLE. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 1

DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2019, THE PATIENT UNDERWENT OPEN REDUCTION INTERNAL FIXATION SURGERY FOR FEMORAL NECK FRACTURE WITH THE FEMORAL NECK SYSTEM (FNS) IN QUESTION. THERE WAS NO SURGICAL DELAY. THE HOSPITAL FOLLOWED UP WITH THE PATIENT, AND THE HOSPITAL CONFIRMED THAT THE NON-UNION OCCURRED AND THE CUTOUT PROGRESSED GRADUALLY. ON (B)(6) 2019, THE PATIENT WILL UNDERGO REMOVAL SURGERY OF THE FNS FOLLOWED BY A BIPOLAR HEMI ARTHROPLASTY (BHA) SURGERY. THE SURGEON COMMENTED THAT THERE WAS NO PRODUCT DEFECT. THE FRACTURE TYPE WAS GARDEN 1. NO FURTHER INFORMATION IS AVAILABLE. THIS COMPLAINT INVOLVED 4 (FOUR) DEVICES. THIS REPORT IS FOR ONE (1) BOLT FOR FEMORAL NECK SYSTEM. THIS IS REPORT 2 OF 4 FOR (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1024057 BOLT F/FEM NECK SYST F/CONSTRUCT LENGTH APPLIANCE,FIXATION,NAIL KTT OBERDORF SYNTHES PRODUKTIONS GMBH L873526

Patients

Seq Age Sex Outcome Treatment
1 89 YR Required Intervention