FDA Adverse Event Malfunction Summary report: N

UNKNOWN (SURPASS STREAMLINE)

MDR report key: 9226591 · Received October 23, 2019

Report

Report Number
3008881809-2019-00320
Event Type
Malfunction
Date Received
October 23, 2019
Date of Event
October 3, 2019
Report Date
December 16, 2019
Manufacturer
STRYKER NEUROVASCULAR CORK
Product Code
OUT
PMA / PMN Number
P170024
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NL
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 0

PRODUCT LONG DESCRIPTION: UNKNOWN (SURPASS STREAMLINE). THE ELECTRONIC DEVICE HISTORY RECORD (EDHR) REVIEW CONFIRMS THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE REPORTED EVENT IS COVERED IN THE DIRECTION FOR USE (DFU). AS WELL, THE RISK OF THE REPORTED EVENT IS DOCUMENTED IN THE RISK DOCUMENTATION AND THERE ARE CURRENT CONTROLS TO MITIGATE THE RISK OF THE AS REPORTED EVENT. THE DEVICE WAS RETURNED AND THE STENT DELIVERY WIRE (SDW) WAS KINKED. THE FLOW DIVERTER WAS FOUND TO BE PREMATURELY DEPLOYED WITHIN THE RETURNED XT-27 MICROCATHETER, DEFORMATION WAS ALSO NOTED TO THE FLOW DIVERTER. IT IS PROBABLE THAT THE DEVICE WAS DAMAGED DURING THE CLINICAL PROCEDURE CAUSING THE REPORTED EVENT. AN ASSIGNABLE CAUSE OF PROCEDURAL FACTORS WAS ASSIGNED TO THE REPORTED ISSUE STENT FAILED/ UNABLE TO OPEN AND TO THE ANALYZED ISSUES SDW KINKED/ BENT, STENT DEPLOYED PREMATURELY DURING USE AND STENT DEFORMED, AS THE ISSUE IS ASSOCIATED WITH A PRODUCT THAT MEETS STRYKER DESIGN AND MANUFACTURE SPECIFICATIONS AND WAS USED IN ACCORDING WITH THE DFU BUT DUE TO PROCEDURAL AND/OR ANATOMICAL FACTORS DURING USE, THE PRODUCT PERFORMANCE WAS LIMITED. THE NEUROVASCULAR STRYKER SURPASS EVOLVE DEVICE IS NOT CURRENTLY NOT APPROVED OR COMMERCIALLY SOLD IN THE USA. THE EVENT FOR THE SURPASS EVOLVE DEVICE WAS FILED TO FDA AS A SIMILAR PRODUCT TO STRYKER DEVICE SURPASS STREAMLINE DEVICE THAT IS COMMERCIALLY AVAILABLE IN THE US (PMA # P170024). (B)(4).

Description of Event or Problem · 0

IT WAS REPORTED THAT DURING THE PROCEDURE, THE SUBJECT STENT DID NOT OPEN; THEREFORE, THE PHYSICIAN REMOVED THE STENT OUT OF THE PATIENT¿S ANATOMY. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH ANOTHER STRYKER DEVICE. NO PATIENT CONSEQUENCES REPORTED DUE TO THIS EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING THE PROCEDURE, THE SUBJECT STENT DID NOT OPEN; THEREFORE, THE PHYSICIAN REMOVED THE STENT OUT OF THE PATIENT¿S ANATOMY. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH ANOTHER STRYKER DEVICE. NO PATIENT CONSEQUENCES REPORTED DUE TO THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1018859 UNKNOWN (SURPASS STREAMLINE) INTRACRANIAL ANEURYSM FLOW DIVERTER OUT STRYKER NEUROVASCULAR CORK UNKNOWN

Patients

Seq Age Sex Outcome Treatment
1 XT-27 MICROCATHETER (STRYKER)