FDA Adverse Event Malfunction Summary report: N

LIFEPAK CR(R) PLUS DEFIBRILLATOR

MDR report key: 922537 · Received October 2, 2007

Report

Report Number
3015876-2007-00375
Event Type
Malfunction
Date Received
October 2, 2007
Report Date
September 5, 2007
Manufacturer
PHYSIO-CONTROL, INC.
Product Code
MKJ
PMA / PMN Number
K011144
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

PHYSIO-CONTROL, INC. EVALUATED THE DEVICE. THE ROOT CAUSE WAS DETERMINED TO BE DUE TO A FAILURE OF THE CAPACITOR, C177 ON THE ANALOG PCB ASSEMBLY. AFTER REPLACING THE ANALOG PCB ASSEMBLY, PROPER OPERATION WAS CONFIRMED AND THE DEVICE WAS RETURNED TO THE CUSTOMER.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE WOULD NOT OPERATE ON DC POWER. THERE WAS NO PATIENT USE ASSOCIATED WITH THIS REPORTED EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LIFEPAK CR(R) PLUS DEFIBRILLATOR MKJ PHYSIO-CONTROL, INC. CR PLUS NA

Patients

Seq Age Sex Outcome Treatment
1 NA YR