FDA Adverse Event Other Summary report: N

ACCELERATOR APS

MDR report key: 922263 · Received October 4, 2007

Report

Report Number
1628664-2007-00445
Event Type
Other
Date Received
October 4, 2007
Date of Event
September 11, 2007
Report Date
September 11, 2007
Manufacturer
ABBOTT MANUFACTURING INC.
Product Code
JQW
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE ACCELERATOR APS USER GUIDE DOES NOT PROVIDE SPECIFIC INFORMATION TO ADDRESS THE SCENARIO. IT DOES PROVIDE INFORMATION RELATED TO PROCEDURE THAT REQUIRE THE SYSTEM TO BE IN THE OFF-LINE STATE AND ALSO FOR ADDRESSING WARNINGS AND ERROR MESSAGES THAT MAY BE GENERATED DURING NORMAL OPERATION: CONFIGURATION SYSTEM (ON-LINE/OFF-LINE). PERFORM THE FOLLOWING PROCEDURE TO TURN A WORKCELL MODULE ON-LINE OR OFF-LINE. WARNING: CHANGING THE ON-LINE/OFF-LINE STATUS OF A WORKCELL MODULE OR INSTRUMENT CHANGES THE PROCESSING CONFIGURATION OF THE WORKCELL. ENSURE TUBES IN THE INPUT LANES ARE APPROPRIATE FOR THE CHANGED WORKCELL CONFIGURATION. THE USER GUIDE ALSO PROVIDES INFORMATION REGARDING HOW THE SYSTEM RESPONDS WHEN IN THE OFF-LINE STATE FOR VARIOUS COMPONENTS OF THE SYSTEM. THIS IS AN INITIAL REPORT. AN INVESTIGATION IS IN PROCESS. A FINAL REPORT WILL BE SUBMITTED WHEN THE INVESTIGATION IS COMPLETED.

Description of Event or Problem · 1

AN ABBOTT IN-HOUSE INVESTIGATION ON THE ACCELERATOR AUTOMATED PROCESSING SYSTEMS (APS) INPUT/OUTPUT MODULE (IOM) FOUND THAT SAMPLE PRESENTATION/SAMPLE QUEUE ERRORS ARE NOT CONSISTENTLY REPORTED WHEN THE ANALYZER IS PUT OFF-LINE: IF THE ANALYZER IS PUT OFF-LINE WHILE A SAMPLE PRESENTATION ERROR OCCURS, THE SAMPLE PRESENTATION ERROR IS REPORTED ONLY AFTER THE ANALYZER IS PUT ON-LINE AGAIN, WHILE THE SAMPLE QUEUE ERROR IS NOT REPORTED. IF THE ANALYZER IS PUT OFF-LINE WHILE THE ANALYZER IS SAMPLING (WITHOUT ANY SAMPLE PRESENTATION ERRORS), THE APS REPORTS A FALSE SAMPLE PRESENTATION ERROR WHEN THE ANALYZER IS PUT BACK ON-LINE. THESE SCENARIOS PROVIDE THE POSSIBILITY THAT THE TESTS ORDERED FOR ONE SAMPLE MAY BE ASPIRATED FROM A DIFFERENT SAMPLE WITHOUT ANY ERROR BEING GENERATED. ALSO, THE POSSIBILITY EXISTS FOR SAMPLE CONTAMINATION SINCE THE PROBE MAY CARRYOVER SAMPLE FROM ONE TUBE TO THE NEXT. TO DATE, THERE HAS BEEN NO IMPACT TO PATIENT MANAGEMENT REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ACCELERATOR APS AUTOMATED SAMPLE HANDLING AND PROCESSING SYSTEM JQW ABBOTT MANUFACTURING INC. NA NA

Patients

Seq Age Sex Outcome Treatment
1 NI YR